Power off the instrument

 

Insert the manual lid opening tool into the hole on the left hand side of the instrument.

 

Push the tool inward, holding it horizontal until the lid lock releases and the lid can be opened.

WARNING: Before opening the lid manually, the rotor must be completely stopped. (Observe the rotor while

opening the lid carefully by hand; if the rotor is still rotating, close the lid and wait approximately 10

minutes before repeating the operation.) (Und das Loch im Deckel? Kann man das nicht dort sehen?)

3

 

Specifications

Technical Specifications

HuMax 4K

Maximum speed

5000 rpm

Maximum RCF

2822xg

Tube capacity

12 x 15 ml

Control system

Programmable Microprocessor Control

Speed set range

1000- 5000 rpm

Speed set step

10 rpm

Timer set range

1- 99 minutes and hold position

Timer set step

1 minute

Motor

Induction Motor

Supply Values

230 V / 50 Hz

Power Consumption

350 W

External Dimensions (WxDxH) mm

274 x 355 x 238

Net / packed weight (kg)

11 / 12

4

 

Accessories

Cat. No.

Tube Type

Max. Tube Diameter (mm)

17155

1.5 / 2 ml microtubes

11

17156

5  ml vacuum / non-vacumm tubes

13

17157

7 ml vacuum / non-vacuum tubes

13

 

6DIHW\,QWHUORFN6\VWHP

The HuMax 4K Centrifuge is equipped with an interlock system that prevents the centrifuge lid from opening when

the rotor is spinning.

The centrifuge will not operate until the lid is closed completely. The lid remains closed until the rotor stops

spinning. The run can not be started before the lid is correctly closed. In this case the lid-open indicator will light.

NOTE: If a power failure occurs, manual access to the sample in the centrifuge is possible using the special tool

included with the centrifuge.

Summary of Contents for HUMAX 4K

Page 1: ...Gtl w 3J User Manual Cat No 15210 100...

Page 2: ...i Qduhrhnm Khrs ne sgd L mt k No DATE Rev REVISION DESCRIPTION 1 01 2007 01 First edition...

Page 3: ...ii...

Page 4: ...llation of this product This warranty does not apply to damages incurred in shipment of goods Any damage so incurred shall be reported to the freight carrier for settlement or claim 3 INTENDED USE OF...

Page 5: ...at least potentially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the labo...

Page 6: ...sibly contaminated parts Before the instrument is removed from the laboratory for disposal or servicing it must be decontaminated Decontamination should be performed by authorised well trained personn...

Page 7: ...b...

Page 8: ...lock System 4 5 Instructions for Use 5 5 1 Schematic Drawing 5 5 2 Controls and Indicators 6 5 3 Preparing The Run 6 5 4 Programming 7 5 5 Program Recall 7 5 6 Starting Running and End Of Run 7 5 7 Er...

Page 9: ...does not allow the centrifuge to operate or the rotor to spin if the lid is not closed The lid locking system allows the lid to be opened when the program ends by pressing a single key There are audib...

Page 10: ...balance The central hole of the rotor and the motor shaft must also be kept clean and dry at all times RDGLQJ Each tube insert must be the same weight as the one diametrically opposite for balancing C...

Page 11: ...d set step 10 rpm Timer set range 1 99 minutes and hold position Timer set step 1 minute Motor Induction Motor Supply Values 230 V 50 Hz Power Consumption 350 W External Dimensions WxDxH mm 274 x 355...

Page 12: ...5 10 XPDQ X0D User Manual 5 Instructions for Use 6FKHPDWLF UDZLQJ Figure 1...

Page 13: ...o set the speed and time during programming and to check the RCF value during the operation 5 Pulse Key Causes the motor to increase to the maximum speed The motor will continue to spin at that speed...

Page 14: ...otor stops spinning When the program is completed END is shown on the speed display 1 and the warning beep sounds to indicate to the user that program has ended The warning beep will continue to sound...

Page 15: ...ere hazardous substances including radioactive chemicals are frequently found Always use the appropriate decontamination procedures where the rotor is exposed to these chemicals Examples of commonly u...

Page 16: ...DLO No daily cleaning is required except in the case of accidental glass breakage or a large amount of spillage in the bowl HHNO Clean the bowl and the accessories with a cotton wool pad dipped in 70...

Page 17: ...10 10 8 Electrical Circuit Diagram...

Page 18: ...HUMAN Gesellschaft f r Biochemica und Diagnostica mbH Max Planck Ring 21 65205 Wiesbaden Germany Tel 49 61 22 99 88 0 Fax 49 61 22 99 88 100 e Mail human human de www human de 01 2007 01...