4/10
Power off the instrument
Insert the manual lid opening tool into the hole on the left hand side of the instrument.
Push the tool inward, holding it horizontal until the lid lock releases and the lid can be opened.
WARNING: Before opening the lid manually, the rotor must be completely stopped. (Observe the rotor while
opening the lid carefully by hand; if the rotor is still rotating, close the lid and wait approximately 10
minutes before repeating the operation.) (Und das Loch im Deckel? Kann man das nicht dort sehen?)
3
Specifications
Technical Specifications
HuMax 4K
Maximum speed
5000 rpm
Maximum RCF
2822xg
Tube capacity
12 x 15 ml
Control system
Programmable Microprocessor Control
Speed set range
1000- 5000 rpm
Speed set step
10 rpm
Timer set range
1- 99 minutes and hold position
Timer set step
1 minute
Motor
Induction Motor
Supply Values
230 V / 50 Hz
Power Consumption
350 W
External Dimensions (WxDxH) mm
274 x 355 x 238
Net / packed weight (kg)
11 / 12
4
Accessories
Cat. No.
Tube Type
Max. Tube Diameter (mm)
17155
1.5 / 2 ml microtubes
11
17156
5 ml vacuum / non-vacumm tubes
13
17157
7 ml vacuum / non-vacuum tubes
13
6DIHW\,QWHUORFN6\VWHP
The HuMax 4K Centrifuge is equipped with an interlock system that prevents the centrifuge lid from opening when
the rotor is spinning.
The centrifuge will not operate until the lid is closed completely. The lid remains closed until the rotor stops
spinning. The run can not be started before the lid is correctly closed. In this case the lid-open indicator will light.
NOTE: If a power failure occurs, manual access to the sample in the centrifuge is possible using the special tool
included with the centrifuge.