A control should be performed:
>
If the meter is being used for the first
time or a new lot of test strips is being
used (indicated by a different code
number).
>
If you suspect that the meter or test
strips are not working properly.
>
Any time you see repeated unexpected
blood glucose test results.
>
Any time you drop or bump the meter.
>
You should check system performance
with the HumaSens Glucose control so-
lution regularly or anytime you suspect
that test results are inaccurate.
1.
Insert a test strip into test port. The
meter will switch on automatically.
2.
The code number will appear on the
screen. Make sure the code number is
the same as the one on the test strip vial.
3.
A blood drop symbol flashing on the
screen means the system is ready for
sample loading.
4.
Shake the control solution vial well,
then open the cap and discard the
first three drops. After cleaning the
dispensing tip with a tissue, drop some
control solution onto a clean, non-
absorbent surface.
5.
Bring the sample inlet of the test strip
in contact with the drop of control
solution until the meter beeps. The
screen will display a countdown from
10 seconds.
6.
After counting down, the screen will
display the control solution test result.
Compare the test results to the control
range shown on the test strip vial.
Chapter 2 Meter Setup