Attaching the General purpose sling to hanger bars

2-point hanger bar

3-point hanger bar

4-point hanger bar

Sling with 8 straps: 

Starting from head, the 3 top 

straps on the left attached on left 

side, repeat on the right side. The 

2 leg straps should be attached to 

the hook on each side.

Sling with 8 straps: 

Starting from head, the 3 top 

straps on the left - attach on 

left side, repeat on right side, 

the 2 leg straps (right and left) - 

attach to the leg hook.

Sling with 8 straps: 

Each hook should have 2 straps 

attached. The  

 

bottom 2 leg straps can be cros

-

sed for increased comfort.

Sling with 4 straps: 

Starting from head, the 1 top 

strap on the left attach on left 

side, repeat on right side(1), The 

2 leg straps should be attached to 

the hook on each side.

Sling with 4 straps: 

Starting from head, the 1 top 

strap on the left - attach on left 

side, repeat on right side(1). 

The 2 leg straps (right and left) 

attach to leg hook.

Sling with 4 straps: 

Each hook should have 1 straps 

attached.  The  

 

bottom 2 leg straps can be cros

-

sed for increased comfort.

Care Instructions

Check for wear and tear

Check the sling for wear and tear. It is important 

to check the fabric, the edging, the seams and 

the lifting straps.
Damaged lifting equipment must not be used.

Cleaning

Human Care slings are intended for individual 

use. To avoid infections or transmission of bac

-

teria, it is important to clean the slings properly. 

All slings should be machine washed and dried 

as soon as possible. Slings should always be 

cleaned after use in a wet situation.
Single Patient Use (SPU) slings can not be was

-

hed, they should be discarded.

Always refer to the label of the sling 

before washing.

Periodic inspection

A periodic inspection must be carried out at 

least every 6 months. See Human Care’s peri

-

odic inspection log which can be found on our 

website 

Product lifetime 

The product´s expected lifetime 3 years or 

150-180 times washing, if the product is used 

as intended and maintained according to the 

manufacturer´s instructions, depending on the 

intensity of use and maximum load applied 

during use. If the product label is no longer 

legible, the product should be discarded.

Warranty and support

If you need information or support, please 

contact your Human Care office, 

 

 or distributor. 

Recycling

Once the product has reached the end of its life, 

it should be disposed according to your coun-

try’s rules, especially for the separate collection 

of electrical and electronic products. 
Correct disposal helps prevent negative 

consequences for the environment and human 

health.

Summary of Contents for NSB-103

Page 1: ...User manual Bruksanvisning Manuel de l utilisateur Gebrauchsanweisung General purpose sling NSB 103 NSB 104 NSB 105 NSB 109...

Page 2: ...ompliances and Standards 4 Safety Instructions 4 General Description 5 Technical Specifications 5 Product Label 7 Suspension Alternatives 8 Hanger Bars for Human Care Lifts 9 Compatibility 9 Sling App...

Page 3: ...and ensure you fully understand them before using this product Important You must read the user manual for your device prior to usage Keep this booklet and information handy for future use Product mod...

Page 4: ...olves risks Therefore you must read the user manuals for the lift and the lifting equipment carefully Always make sure that the lifting equipment you are using is intended for use with the lift As a c...

Page 5: ...mm 17 318 kg 700 lbs General purpose sling reusable NSB 103 L Polyester net 8 straps L 1143 mm 45 1029 mm 41 533 mm 21 318 kg 700 lbs General purpose sling reusable NSB 103 XL Polyester net 8 straps X...

Page 6: ...lbs General purpose sling single patient use with clips NSB 104C XL polymethylpentene 4 straps XL 1334 mm 53 1143 mm 45 679 mm 27 318 kg 700 lbs General purpose sling reusable with clips NSB 103C S P...

Page 7: ...ot Bleach c Tumble Dry Low Heat d Do Not Dry Clean e Never Iron f Do Not Use Fabric Softener 18 Owner name For example write the client s name or another unique identifier 19 Indicate annual inspectio...

Page 8: ...section through the strap on the other leg section before attaching them to the hanger bar An individual assessment of the client s needs must always be carried out before choosing an alternative conn...

Page 9: ...er bars made by manufacturers other than Human Care the combined system needs to be designed with the same sling attachment method for the hanger bar of floor lifts and ceiling lifts meaning loop sty...

Page 10: ...aps please see page 2 There are diffe rent ways to attach the loops to a hanger bar and to position the legs This sling can be used both with 2 3 or 4 point hanger bars The image above shows attachmen...

Page 11: ...e the desired positioning of the legs 4 Lower the hanger bar and securely attach straps see page 2 with appropriate loop For sling with clips Attach the leg section lifting clips to the lifting frame...

Page 12: ...achieve the divided position bring the leg sections up between the thighs thread one leg strap through the bottom opening of other left and attach the leg straps to the hanger bar cross over style For...

Page 13: ...the seams and the lifting straps Damaged lifting equipment must not be used Cleaning Human Care slings are intended for individual use To avoid infections or transmission of bac teria it is important...

Page 14: ...Human Care slings are intended to be used with one patient at a time To avoid possible trans mission of infectious diseases it is important to clean the sling if it has become soiled or if it is to b...

Page 15: ...99533 Rev01 ML User Manual General Purpose Sling May2021...

Reviews: