background image

9

ENGLISH

Hanger Bars for Human Care Lifts

Accessory

Description

Part no.

2-point hanger bar

For stationary lift HeliQ. 3 widths:

35 cm/13,8”: 72754

45 cm/17,7”: 72755

60 cm/23,6”: 72757

2-point hanger bar

For portable lifts Roomer S / Altair, 

2 widths:

45 cm/17,7”: 55751

55 cm/21,7”: 55764

4-point hanger bar

For stationary lift HeliQ

70800

4-point suspension

4-point suspension accessory (49 cm). Enables 4-point lifting with the Altair and Roomer S 

lifts. The suspension bar is placed on the lift’s single leg hook.

55619

HeliQ

HeliQ

Roomer S / Altair

(55751 45 cm, 17,7” / 55764 55 cm, 21,7”)

(7275x)

(70800)

Compatibility

Human Care Group slings are designed for use 

with all Human Care lifts and hanger bars. The 

products are also compatible with products of 

other suppliers utilizing the same attachment 

method for securing slings to a hanger bar and 

lift. However, due to the wide range of products 

worldwide, Human Care cannot be responsible 

for any errors that may result from improper 

use or application of combinations of slings, 

hanger bars and lifts of other suppliers. The use 

of the combined system is at the prescriber’s 

own risk and liability. 
To ensure the safety of the client and caregiver, 

Human Care recommends the following mini

-

mum requirements: 
1.  When using slings with lifts and hang

-

er bars, made by manufacturers other 

than Human Care, the combined system 

needs to be designed with the same sling 

attachment method for the hanger bar of 

floor lifts and ceiling lifts, meaning loop 

style slings are combined with a loop style 

hanger bar, clip style slings are combined 

with a clip style hanger bar.

2.  The involved patient lifts and sling must be 

CE marked in accordance with the medical 

device directive, MDD 93/42/

 

EEG / upcoming MDR 2017/75 and comply 

with standard requirements for patient lifts 

as stated in EN ISO 10535.

3.  The caregiver must follow the manufactu

-

rer’s instruction manuals and recommen

-

dations with respect to the use, care, 

client, caregiver, cleaning, and inspection 

of the slings and lifts. 

4.  Different products on the assembled lift 

system: i.e., lift, hanger bar, sling, scale and 

other accessories may have different max. 

loads. It is always the lowest maximum 

allowable load indicated for the respective 

products and the system that applies for 

the entire system. 

 

Always check the markings on all individual 

products in the system.

5.  Individual risk assessment, including a 

practical test of the system and validation 

with the specific combination used are 

required to affirm that the size of the sling 

is correct for the intended use and compa-

tible to the width and design of the hanger 

bar. The combined system needs to be 

documented by the competent assessor. 

If there are any questions, please contact your 

local Human Care office or distributor.

Summary of Contents for NSB-103

Page 1: ...User manual Bruksanvisning Manuel de l utilisateur Gebrauchsanweisung General purpose sling NSB 103 NSB 104 NSB 105 NSB 109...

Page 2: ...ompliances and Standards 4 Safety Instructions 4 General Description 5 Technical Specifications 5 Product Label 7 Suspension Alternatives 8 Hanger Bars for Human Care Lifts 9 Compatibility 9 Sling App...

Page 3: ...and ensure you fully understand them before using this product Important You must read the user manual for your device prior to usage Keep this booklet and information handy for future use Product mod...

Page 4: ...olves risks Therefore you must read the user manuals for the lift and the lifting equipment carefully Always make sure that the lifting equipment you are using is intended for use with the lift As a c...

Page 5: ...mm 17 318 kg 700 lbs General purpose sling reusable NSB 103 L Polyester net 8 straps L 1143 mm 45 1029 mm 41 533 mm 21 318 kg 700 lbs General purpose sling reusable NSB 103 XL Polyester net 8 straps X...

Page 6: ...lbs General purpose sling single patient use with clips NSB 104C XL polymethylpentene 4 straps XL 1334 mm 53 1143 mm 45 679 mm 27 318 kg 700 lbs General purpose sling reusable with clips NSB 103C S P...

Page 7: ...ot Bleach c Tumble Dry Low Heat d Do Not Dry Clean e Never Iron f Do Not Use Fabric Softener 18 Owner name For example write the client s name or another unique identifier 19 Indicate annual inspectio...

Page 8: ...section through the strap on the other leg section before attaching them to the hanger bar An individual assessment of the client s needs must always be carried out before choosing an alternative conn...

Page 9: ...er bars made by manufacturers other than Human Care the combined system needs to be designed with the same sling attachment method for the hanger bar of floor lifts and ceiling lifts meaning loop sty...

Page 10: ...aps please see page 2 There are diffe rent ways to attach the loops to a hanger bar and to position the legs This sling can be used both with 2 3 or 4 point hanger bars The image above shows attachmen...

Page 11: ...e the desired positioning of the legs 4 Lower the hanger bar and securely attach straps see page 2 with appropriate loop For sling with clips Attach the leg section lifting clips to the lifting frame...

Page 12: ...achieve the divided position bring the leg sections up between the thighs thread one leg strap through the bottom opening of other left and attach the leg straps to the hanger bar cross over style For...

Page 13: ...the seams and the lifting straps Damaged lifting equipment must not be used Cleaning Human Care slings are intended for individual use To avoid infections or transmission of bac teria it is important...

Page 14: ...Human Care slings are intended to be used with one patient at a time To avoid possible trans mission of infectious diseases it is important to clean the sling if it has become soiled or if it is to b...

Page 15: ...99533 Rev01 ML User Manual General Purpose Sling May2021...