3.2 Back Panel Elements
[1]
PLUG FOR NON-HEATING APPARATUS (C13) WITH MAINS SWITCH AND
FUSEHOLDER
Fuse 4AT (5x20 mm)
[2]
IN-2 100 KΩ BNC SOCKET
100 kΩ signal input; isolated BNC socket
Reference potential is connected to protective earth via 1 MΩ.
Connect to socket [3] in bridge operation.
[3]
BRIDGE OUT BNC SOCKET
Inverted preamp output for simple setup of bridge circuits
[4]
INTERLOCK 1
BNC socket with shorting plug
Remote-controlled safety shutdown turns off the amplifier via an external switch (breaker,
normally closed contact)
If safety shutdown was triggered: AMPLIFIER ON LED is OFF and READY LED is ON
[5]
INTERLOCK 2
BNC socket with shorting plug
Remote-controlled safety shutdown turns off the amplifier via an external switch (breaker,
normally closed contact)
A1230-02