Properties of standard variables   

TECHNICAL BULLETIN 

 

Chapter 6 

 

Profibus® Gateway SE 

Liability for errors and misprints excluded. 

V1.2.0en/05.02.16   

10 

6.7

 

“vStatus1” – Status of the thermostat 

Query current thermostat status. 

The returned value is a bit field. The individual bits must be considered independently of each other. 

 

Bit 

Meaning 

Explanation 

“1” 

“0” 

0  Temperature control operating mode 

active 

Inactive 

1  Circulation operating mode  

active 

Inactive 

2  Refrigerator compressor 

enabled 

disabled 

3  Temperature control mode “Process control” 

active 

Inactive 

4  Circulating pump 

enabled 

disabled 

5  Cooling power available 

available 

unavailable 

6  Tkeylock 

active 

Inactive 

7  PID parameter set, temperature controller 

automatic mode 

expert mode 

8  Error 

fault occurred 

no fault 

9  Warning 

new warning  

received 

no new warning 

10  Mode for presetting the internal temperature   

active 

inactive 

11  Mode for presetting the external temperature 

active 

inactive 

12  DV E-grade 

activated 

not activated 

14  Restart electronics / Power failure (*) 

no restart 

restart 

15  Freeze protection (not available on all devices) 

active 

inactive 

(*) This bit is reset to 0 when the control is restarted. If the variable vStatus1 is read for the first time after a 

restart, this bit has not been set. In all further reading procedures, this bit will send back a 1. In this way, an 

unexpected restart can be recognized. 

6.8

 

“vTmpMode” – Temperature control mode 

Set and query the temperature control mode of the thermostat. 

0: Temperature control mode internal 

1: Temperature control mode process (cascade control) 

6.9

 

“vTmpActive” – Temperature control 

Start or stop temperature control of the thermostat or query current status. 

0: Temperature control not active 

1: Temperature control active 

6.10

 

“vSNRL + vSNRH” – Serial number 

Query the serial number of the thermostat. 

Because the serial number can have larger values than 65535 but only 2 bytes per variable are available, 

the serial number is made up of 2 data words (Low + High) each with 2 bytes into a 4 byte value. 

Variable 0x1B (vSNRL) contains the low value bytes, Variable 0x1C (vSNRH) contains the high value bytes. 

“vStatus1” 

Summary of Contents for Profibus SE

Page 1: ...Technical Bulletin Profibus Gateway SE V1 2 0...

Page 2: ...TECHNICAL BULLETIN Profibus Gateway SE Liability for errors and misprints excluded V1 2 0en 05 02 16 2...

Page 3: ...TECHNICAL BULLETIN V1 2 0en 05 02 16 Liability for errors and misprints excluded Profibus Gateway SE 3 Profibus Gateway SE...

Page 4: ...TECHNICAL BULLETIN Profibus Gateway SE Liability for errors and misprints excluded V1 2 0en 05 02 16 4...

Page 5: ...into the process automation system 8 5 1 Basic mode of operation of the Profibus Gateway SE 8 5 2 Description of the standard variant 8 6 Properties of standard variables 9 6 1 vSP Setpoint temperatur...

Page 6: ...standard version presented is designed to cover a large number of process requirements We can offer a customized version if you have addi tional requirements Please note the inherent restrictions of E...

Page 7: ...topology in accordance with the requirements of the PROFIBUS user organization A bus termination can be enabled on the Pro fibus Gateway SE The Profibus Gateway SE has been built and qualified in acc...

Page 8: ...fore always available when retrieving the process image The temperature control unit can immediately return the current process values when a Profibus frame is received On the other hand the update lo...

Page 9: ...lute Absolute pump pressure at the pressure sensor in the thermal fluid circuit of the thermostat only if there is a pressure sensor 6 4 vError Error report The number of the error report that appeare...

Page 10: ...e activated not activated 14 Restart electronics Power failure no restart restart 15 Freeze protection not available on all devices active inactive This bit is reset to 0 when the control is restarted...