5. Test Operation
1.Manual boost mode
After starting up, enter the main menu
(
P 4-4.1), rotate the control lever to make the cursor
point to the "manual boost" option, click the control lever and enter the parameter setting
interface of manual test mode
(
P 4-4.3). After setting various parameters according to
specific test requirements, enter the test boost interface. At this point, press the "high
voltage on" button to switch on the high voltage loop, and then rotate the control lever
(press the control lever to switch the voltage and adjust the fineness). When the voltage
rises to the set test voltage, the timer will start to time, and the system will automatically
drop the voltage back to zero after the timing, and then a record interface of test results
will show up, prompting to save or print, and after making the selection, the test will be
finished. If the voltage does not reach the set test voltage value, the 0.75U
DC1mA
button is
pressed, the system will automatically reduce the voltage to 0.75 times of the current
voltage, and then countdown to the set pressurization time. After the timing is completed,
the system will automatically reduce the voltage to zero, and then pop up the interface of
test results.
2.Automatic boost mode
After starting up, enter the main menu, rotate the control lever to point to the option of
"automatic boost", click the control lever to enter the interface of automatic mode
parameter setting
(
P 4-4.4). After setting various parameters according to specific test
requirements, enter the interface of node setting. At this time, after pressing the “HV on”
button, the system will start to automatically boost the voltage. After rising to the set test
voltage, it will be stabilized and timing will start, and the value of each node will be
extracted. After the timing is completed, the system will automatically reduce the voltage
to zero.
3.MOA automatic test mode
After starting up, enter the main menu, rotate the control lever to point to the "MOA
boost" option, click the control lever and enter the automatic mode parameter
setting
interface
(
P 4-4.5). After setting various parameters according
to
specific
test