automatically don’t need manual click.
Pole quantity: 2/4/6/8
Test mode:
1) Test on pole: Just test pole internal resistance.
2) Test the connection strip: Test pole internal resistance then test the resistance of
pole connection strip.
Test item
1)
Test voltage and internal resistance together.
2)
Just test voltage no internal resistance.
3)
Test all voltage firstly then test internal resistance.
It’s unnecessary to set float charging voltage, float charging current, battery
temperature won’t affect test.
Start test: After setting specifications the clip instrument test clips on battery and click
“Start test”.
Return: Return to last interface.
3.2.2 Existing battery specification
Main interface”
” —“
”, choose setting parameters
Fig 3.2.2 Existing battery specification
a
、
“Delete all” : Click it will delete all modular record.
b
、
“Delete”: Select certain record and then click it will delete present record.
c
、
Last page, next page: Find out record.
d
、
Check setting: Select certain record click “ check setting” enter the specification
setting board.