5
implemented in accordance with relevant national standards and plans:
GB/T16927-1996 High-Voltage Experimentation Technology and DL/T596-1996
Preventive Testing Regulations for Electrical Equipment.
2.
Due to the design characteristics and structure limitation of this series products,
continuous operation under rated output capacity should not exceed 2 hours. If the
products need to be used again, the power can be connected until power off for 2
hours. Continuous operation can be done under two thirds of rated voltage and two
thirds of rated current.
3.
In normal insulation of high voltage appliance, this series products are not allowed to
operate in case of exceeding rated output voltage.
4.
When this series products consist of cascade high voltage test, users should pay
special attention to check the correctness of conditions and wiring of Class 2 and 3
transformer's main body and insulating bracket, as well as the protective grounding
of entire system and conditions of protective grounding.