13

complete, the instrument will automatically print the measuring results (if you choose the

automatic printing function).

VIII.Use the Matters Needing Attention

1, test the cup must be cleaned with anhydrous ethanol, stay dry rear can use.

2, pay attention to the determination of acid value before, must first blank experiment, if a

measuring cup sample can make a blank experiment. If the time interval is longer or the

environment temperature changes, the each measured sample before to do an blank test.

Please confirm and check whether it is the first cup of cup into the stir bar magnet.

3, special remind: blank test, neutralization liquid drop speed must not be too fast, or you

will make the drop testing is not accurate, affect the experiment result.

4, if a blank test, open cover, the sample inspection should also uncovered protective

cover, when the ambient light is consistent before and after the test.

5, experimental oil cup cup dish, the pallets must ensure that the bottom is completely

inside the cup, in order to avoid cup set is not straight, effect light detection.

6, every cup after the experiment must be removed from the cup dish, to avoid impact

position when doing the experiment again.

7, at the end of the experiment, and fluid and extraction fluid bottle should be placed on

shading in sealed preservation, avoid deterioration failure.

IX.Failure Phenomenon and Elimination Method

1, the neutralization liquid droplet, slow or little.

1) Possible and liquid bottle cap not tighten, air leakage. Tighten the lid.

2) Likely due to not completely clean, make the neutralization titration needle

clogging

Processing method:

Replace # 4.5 needles, adjust the tilt Angle of the needle, the droplet number detection

accuracy.

2, low acid value oil test results appear error 0.0001 mgKOH/g or phenomenon.

Solution: need to blank experiment. To ensure that the blank experiment must be

accurate.

Summary of Contents for HZCS-X6

Page 1: ...HZCS X6 Automatic Oil Acidity Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...nical Parameters 2 IV Instrument Structure 3 V Menus and Buttons Operation Instructions 4 VI Out of The Installation 12 VII Test Steps 12 VIII Use the Matters Needing Attention 13 IX Failure Phenomeno...

Page 3: ...uitable for the detection of transformer oil turbine oil and fuel resistance acid value of the sample analysis of measurement The product through mechanical optical and electronic technology integrate...

Page 4: ...n and CO2 in the air Users don t have to hand contact with solvents reducing the organic solvents and chemicals to the harm of human body 9 the equipment is simple operation using large screen LCD and...

Page 5: ...ent is divided into three parts namely control box mechanical transmission parts test area Test area is placed in the oil cup wheel number of drops of detection device automatic determination of acid...

Page 6: ...ument becomes simple Instrument is divided into four display welcome boot images parameter Settings screen history picture sample pictures Detailed below 1 the boot screen Connect good instrument powe...

Page 7: ...nt factory has been calibrated no special circumstances do not easily into the calibration Remember that If the boot without touching the cross cursor after 2 seconds automatically enter the boot welc...

Page 8: ...e current parameter values press save to save parameters press exit button to return to the boot screen The meaning of each parameter stated below Date time instrument of the system clock Print mode i...

Page 9: ...minimum record number is 1 Print print the current history records the printing content including 1 sample to 6 of the sample determination results the inspection date and time Del delete the current...

Page 10: ...ation national standard and concentration in 0 02 to 0 05 moL L if and the concentration of the night and showed inconsistent by neutralizing night concentration increase and decrease the Numbers belo...

Page 11: ...experiment process with sample determination 3 Cur Hue is value of the color sensor automatic detection and display instrument 4 End Hue experimental basis of titration end point is refers to the test...

Page 12: ...Note if you quit is not to continue the line experiments please cleaning operations or you can press SKIP exit Sample test used to determine the acid value of crude oil sample After open the oil cup...

Page 13: ...on to adjust parameters as shown in the figure below After all the input press the RUN key are measured and the process used in the determination of the Test result column indicates display informatio...

Page 14: ...e will automatically blank experiment operation note to observe the neutralization liquid titration speed about every drop 1 seconds and then will finish cup to take out the blank experiment experimen...

Page 15: ...when the ambient light is consistent before and after the test 5 experimental oil cup cup dish the pallets must ensure that the bottom is completely inside the cup in order to avoid cup set is not st...

Page 16: ...rinter paper began to walk After the front end of the printing paper cut into triangles printing paper into the printer back into paper printing paper will be involved in printer Typed to stay out of...

Reviews: