background image

12

VI.Out of The Installation

1, when unpacking the case firmly laid instrument laboratory work platform.

2, take a beaker in random with 80 ml water, potassium hydroxide dissolves adequately,

then inject gas bottle.

3, to extract randomly with the bottle injection extract (liquid level shall not be higher than

500 ml scribed line).

4, open the back panel, the extracted liquid bottle and the neutralization liquid bottle into

the fixed bottle hole at the back of the instrument.

5, on 220 v power supply. Note: 220 v power supply should be good grounding line!

VII.Test Steps

1, open the power supply, instrument display page.

2, experiments: blank, if change the extract, blank experiment must do first before sample

test. Get empty cup into one cup, and stir into a grain. Adjust the sample quality is "blank",

press "start" button, the machine will automatically blank experiment operation. (note: to

observe the neutralization liquid titration speed about every drop 1 seconds) and then will

finish cup to take out the blank experiment, experiment.

3, sample test:

1) press "set." key in the boot screen, enter the parameter setup screen, enter the

neutralization liquid concentration.

2) with finger touch to modify parameters are selected, press "add", "decrease" button

input neutralization liquid concentration.

3) press the "RUN" key in the boot screen, enter the sample picture.

4) sample weighing: sample with a syringe, insulating oil, and the quantity of turbine oil is

8 ml, fire-resistant oil to 1 ml, said its net weight number and record a cup.

5) will the oil cup into the cup to the number plate, respectively, in a stirring.

6) with your fingers touch the selected experimental cup corresponds to the sample

quality, and quality of input samples.

7) press the "RUN" key to determine sample automatically.

8) every finished a sample, the screen shows the test results, for the sample set is

Summary of Contents for HZCS-X6

Page 1: ...HZCS X6 Automatic Oil Acidity Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...nical Parameters 2 IV Instrument Structure 3 V Menus and Buttons Operation Instructions 4 VI Out of The Installation 12 VII Test Steps 12 VIII Use the Matters Needing Attention 13 IX Failure Phenomeno...

Page 3: ...uitable for the detection of transformer oil turbine oil and fuel resistance acid value of the sample analysis of measurement The product through mechanical optical and electronic technology integrate...

Page 4: ...n and CO2 in the air Users don t have to hand contact with solvents reducing the organic solvents and chemicals to the harm of human body 9 the equipment is simple operation using large screen LCD and...

Page 5: ...ent is divided into three parts namely control box mechanical transmission parts test area Test area is placed in the oil cup wheel number of drops of detection device automatic determination of acid...

Page 6: ...ument becomes simple Instrument is divided into four display welcome boot images parameter Settings screen history picture sample pictures Detailed below 1 the boot screen Connect good instrument powe...

Page 7: ...nt factory has been calibrated no special circumstances do not easily into the calibration Remember that If the boot without touching the cross cursor after 2 seconds automatically enter the boot welc...

Page 8: ...e current parameter values press save to save parameters press exit button to return to the boot screen The meaning of each parameter stated below Date time instrument of the system clock Print mode i...

Page 9: ...minimum record number is 1 Print print the current history records the printing content including 1 sample to 6 of the sample determination results the inspection date and time Del delete the current...

Page 10: ...ation national standard and concentration in 0 02 to 0 05 moL L if and the concentration of the night and showed inconsistent by neutralizing night concentration increase and decrease the Numbers belo...

Page 11: ...experiment process with sample determination 3 Cur Hue is value of the color sensor automatic detection and display instrument 4 End Hue experimental basis of titration end point is refers to the test...

Page 12: ...Note if you quit is not to continue the line experiments please cleaning operations or you can press SKIP exit Sample test used to determine the acid value of crude oil sample After open the oil cup...

Page 13: ...on to adjust parameters as shown in the figure below After all the input press the RUN key are measured and the process used in the determination of the Test result column indicates display informatio...

Page 14: ...e will automatically blank experiment operation note to observe the neutralization liquid titration speed about every drop 1 seconds and then will finish cup to take out the blank experiment experimen...

Page 15: ...when the ambient light is consistent before and after the test 5 experimental oil cup cup dish the pallets must ensure that the bottom is completely inside the cup in order to avoid cup set is not st...

Page 16: ...rinter paper began to walk After the front end of the printing paper cut into triangles printing paper into the printer back into paper printing paper will be involved in printer Typed to stay out of...

Reviews: