15
FIG 5-13 No data screen
VI. Precautions And Instrument Maintenance
1) The instrument should be used in a non-corrosive environment.
2) The oil cup should be cleaned, when replacing test sample.
3) Check thermocouple should not have oil contamination, or it will affect the detection
sensitivity.
4) No touch ignition head, oil cup and its surrounding parts to avoid scorching during the
test.
5
)
For instruments that use electric ignition, the electric ignition head needs to be replaced
regularly.
VII. Failure And Treatment
NO.
Failure
Treatment
1
Turn on the power,
the LCD screen no
display
Check power supply is plugged in, open the instrument measured
the board examine the plug is not loose.