12
1. Press the cleaning time to set the time required for cleaning. Enter the time on the
numeric keypad of the keypad in seconds. Press the key to confirm the entry and the
key to cancel. The original value is 120 seconds.
2. Set the analysis time according to the analysis time. Enter the time in seconds on the
numeric keypad. Press the confirm key on the keypad to confirm the entry and the
cancel key to cancel. The original value is 60 seconds.
3. Press the oil drain time to set the time required for oil drain. Enter the time on the
numeric keypad of the keypad in seconds. Press the key to confirm the entry and the
key to cancel. The original value is 18 seconds.
4. Press the oil filter time to set the time required for draining oil. Use the numeric
keypad to enter the time in seconds. Press the enter key to confirm the entry and the
cancel key to cancel. The original value is 180 seconds.
5. Set the oil sample volume according to the oil sample volume. Enter the oil sample
volume in mL using the numeric keys on the keypad. Press the OK key on the keypad
to confirm the entry and the Cancel key to cancel. The original value is 25 mL.
6. Press the calibration key to execute the calibration procedure.
7. Press the verification key to execute the verification procedure.
8. Press the drain button to execute the drain procedure.
9. Press the write parameter key to save the setting input.
10. Push
↖
to return to the main screen.
1
)、
Calibration: Press the calibration button to enter the calibration program, the
instrument will automatically quantify 25ml of oil, and after the oil filter is prompted
to use a 1ml syringe to inject air into the degassing tank 0.15, 0.35, 0.5, 0.5, 0.5,
0.5ml, It is used to find a calibration curve within 10% of the gas content in the oil.
When making an oil sample, the gas content in the oil is calculated based on this
curve. Note that the instrument has been calibrated in detail before it leaves the
factory. Generally, calibration is not required during use. If it is necessary to
recalibrate, you must choose a syringe with a good seal, and immerse one end of
the needle core with some transformer oil before inserting it into the needle tube. It
is best that the joint between the core and the needle tube is filled with oil. This oil