6

7. Vacuum degree display: used to display the true value of the detected vacuum degree.

(The reading is taken by scientific notation)

8. Pa (vacuum) light: used to indicate the current reading type and unit

9. A (current) light: used to indicate the current reading type and unit. For details, see

3.3.4 Operating Instructions.

10. High-voltage selector switch: used to select the high-voltage level required for

measurement. The tube type to be tested is the high-end selected above 10KV. Choose

low gear below 10KV.

11. Magnetron voltage: used to indicate the voltage of the magnetron coil.

12. Key: the selection key is used to select the tube type, and the light of the selection

key is on to be effective; after the tube type is selected, the light of the selection key is

off to be selected.

Up and down keys: used to scroll up and down to select the desired tube type.

Print key: After the measurement is over, print out the measured data. The print button

light is on and effective.

Check key: If there is no alternative parameter, you can use the check key. This can

directly produce ionization current. Generally speaking, the ionization current (A) is

about two orders of magnitude smaller than the vacuum degree (Pa).

Measurement key: After the tube type is selected, the measurement key light is on to

indicate the start of measurement.

Reset button: used to reset the whole machine after measurement. Press

Reset key

,

the instrument returns to the initial state. If selected during measurement

If the selection is wrong, you can press [Reset Key] at any time to re-select.

13. Printer: used to print test data. The S/L key is the paper feed and eject key; the POW

is the power indicator. After the printing paper is used up, press the buckles on both

sides of the upper cover, open the upper cover, take out the paper roll, and load new

paper.

14. Connect to the earth: used to connect with the earth to reduce the influence of static

electricity.

Summary of Contents for HZ-ZKD

Page 1: ...HZ ZKD Vacuum switch vacuum tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrum...

Page 3: ...ntents I Overview II Key Function And Feature III Major Technical Indicators IV Operational Approach V Precautions VI Maintenance VII Oil Cup Cleaning Method and Common Fault Clearances IX Packing Lis...

Page 4: ...grid voltage matches the power supply voltage of the machine AC 220 10 50 60HZ 7 Please be careful not to use the machine under thunderstorm conditions 8 Please do not damage the cables that come wit...

Page 5: ...vacuum degree in the vacuum interrupter has a very accurate correspondence with the amount of ionized charge The peak value of ionization current is only indirectly related to probability Therefore o...

Page 6: ...errupter of the order of 10E 4Pa We adopt the method of secondary starting high voltage deducting the leakage current and guarantee the vacuum measurement accuracy of the vacuum interrupter whether in...

Page 7: ...ollide with residual gas molecules to ionize The ion current generated is approximately proportional to the residual gas density that is the degree of vacuum For vacuum tubes with different diameters...

Page 8: ...put the magnetron current required for the test to accelerate the electron flow to reduce the influence of various external factors on the measurement accuracy 3 High voltage output terminal used to o...

Page 9: ...rint key After the measurement is over print out the measured data The print button light is on and effective Check key If there is no alternative parameter you can use the check key This can directly...

Page 10: ...the on position and the word P appears in the first digit of the digital display After verification step 3 2 is normal the instrument is operating normally and it can be started after 5 minutes of wa...

Page 11: ...ubble tube type into the tube type of the instrument display window If the tube type to be measured is not indicated in the manual the diameter outer diameter of the vacuum bubble can be measured on s...

Page 12: ...ean because the surface is dirty which may cause the leakage current to change value greater than the ionization current value In this way the measured value after subtracting the leakage current is l...

Reviews: