background image

8

In Figure 5, when the cursor is in the data query, press the select key to scroll up, and

press the start key to scroll down. Move the cursor to print, press the select key according

to the prompts to print the currently displayed data, move the cursor to the U disk for

storage, and press the select key according to the prompts to save the current displayed

data to the USB.

. Wiring method

Note: All connecting surfaces of the test clamp should be in reliable contact with the

sample. If oxidation is found on the contact surface, the oxide layer on the contact surface

should be removed first.

1. Wiring method: As shown in Figure 6, connect the dedicated test wire to the

corresponding I+ and I- terminals according to the color red to red, black to black, and

tighten the thin power wire into V+, In the V- socket, two clamps clamp the two ends of

the tested product.

2. Measurement: After completing the wiring in Figure 6, turn on the power switch and

select the test time for testing. The instrument will automatically complete charging,

steady current and resistance measurements.

.Safety measures

1. The operator should have common sense in the use of general electrical equipment or

instruments.

2. After the test, press the reset button to turn off the power switch. Note: It is strictly

forbidden to disassemble and assemble the test cables when powered.

Summary of Contents for HZ-5300

Page 1: ...HZ 5300 Contact Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrum...

Page 3: ...what will you notice before testing III Brief Introduction Working theory V Technical parameter VI PLC interface VII Operation Instructions and Procedures VIII Applying Illustration IX Calibration X...

Page 4: ...artments Claim The output current and measuring resistance are simultaneously output on the 65K true color LCD and the measurement results can be voltage output or stored in a USB flash drive The powe...

Page 5: ...lution min 0 1 Accuracy 0 5 2 words Power 1000W Working mode continuous measurement Power supply AC220V 10 50HZ Operating temperature 0 40 Relative humidity less than or equal to 90 without condensati...

Page 6: ...Press this key to select various settings of the instrument and the data can be printed after measurement 9 Circulation key main menu cycle selection key or U disk storage 10 Confirm key Press the con...

Page 7: ...ropriate measurement time and measurement current press the start button to start the current the screen displays charging a few seconds later it displays testing and a few seconds later it displays t...

Page 8: ...to move the cursor to the clock to modify press the select key to enter the interface as shown in Figure 4 Figure 4 Press the cycle key to move the cursor between each date data press the select key...

Page 9: ...uld be removed first 1 Wiring method As shown in Figure 6 connect the dedicated test wire to the corresponding I and I terminals according to the color red to red black to black and tighten the thin p...

Page 10: ...sed to measure loop resistance and it is not allowed to measure inductive loops After sales service The instrument will be repaired and replaced free of charge for product quality problems within one...

Reviews: