background image

1

I.Important Hint

If the instrument is not in use, please turn off the power in time.

If the instrument is not used for a long time, please charge and discharge regularly.

Batteries should be charged and discharged at least once a month.

It is strictly forbidden to use power shortage, which will severely shorten the battery

life and even make the battery scrap. When the instrument is short of power, the power

supply should be switched off and charged immediately. Avoid battery failure due to

excessive battery discharge time.

Charging lamp: the charging lamp on the charger is bright red during charging and

bright green after charging.

Users must not disassemble the instrument and replace the battery without authorization.

When the instrument or battery fails, please return to the factory.

II. Introduction

Portable DC Winding Resistance Tester is an innovative product with compact size,

hand-held operation, more portable and easy to carry.

The product is not only suitable for the measurement of transformer, mutual inductor,

reactor and other inductive test products, but also suitable for the measurement of copper

bars, conductors, switch contacts and other resistive test products. The instrument has

fast testing speed and high accuracy.

III. Packaging Content

After receiving the packing box, open the packing box and check if there is any

damage. If the freight packing box is damaged or the liner material is indented, please

inform the freight company or our salesman immediately.

Please check if you have received the following items in the product package

Standard accessories

√1 Hand-held transformer winding resistance tester

√1 sets of test lines (yellow, green, red and black lines each)

√1 external printer (with printing cable attached

Summary of Contents for HZ-3310E-I

Page 1: ...HZ 3310E I Portable DC Winding Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd Dear user...

Page 2: ...l and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instr...

Page 3: ...aging Content 1 IV Functional Characteristics 2 V Technical Indicators 3 VI Adjust the Wristband 3 VII Battery Charging 4 VIII Tilt Hand held Tester 4 IX Product Appearance 4 X Operation Instructions...

Page 4: ...return to the factory II Introduction Portable DC Winding Resistance Tester is an innovative product with compact size hand held operation more portable and easy to carry The product is not only suita...

Page 5: ...tically selected which is convenient and fast Wide range and high accuracy 500u 50K Three phase automatic testing can be carried out and the three phase unbalance rate can be calculated automatically...

Page 6: ...and appearance Working power supply Built in lithium battery or external charger charger input 100 240VAC 50HZ 60HZ Charging voltage 16 8V Charging current 3A Charging time About 4 hr Use time More t...

Page 7: ...east 2 hours and continue to use the hand held tester when charging When the battery is fully charged the charger light changes from red to green VIII Tilt Hand held Tester In order to take the instru...

Page 8: ...5 Front view Side view Display panel Key Wristband...

Page 9: ...on the current operation Power switch The power switch of the whole machine is switched to the open position when it is switched on Switch to the close position when it is switched off Function module...

Page 10: ...icator of the printer are integrated When the printer is powered on the indicator light is normally on and it flashes when the paper is missing Press the button once and the printer passes the paper P...

Page 11: ...is low No Instruction 4 The main menu operation area of the instrument select corresponding functions through keys and press Enter to enter the corresponding menu DC resistance test function can measu...

Page 12: ...current temperature of the tested product and the temperature value is from 99 to 99 The device can monitor the oil temperature inside the test product in real time through an external wireless temper...

Page 13: ...t the appropriate function press the Enter button to execute the currently selected function and press the Cancel button to return to the previous screen No Instruction Retest according to the current...

Page 14: ...the power is not fully discharged It is not allowed to dismantle the test line and cut off the power switch during the test or discharge process In the process of testing transformers the unmeasured...

Reviews: