9
No.
Instruction
1
First-level operation directory, through the “↑↓” key to select these
functions, when these functions are selected, press the “←”→”" key
to select the parameters of the corresponding functions.
Tip: The cursor can quickly jump to the button by pressing the
"Enter" button under the first-level operation directory at “Start Test”.
Testing
temperature
The current temperature of the tested product is set from -99 to +99
C. The device can monitor the samples in real time through an
external wireless temperature probe.
No.
Instruction
Testing
temperature
Internal oil temperature can also be manually set temperature. When
the external wireless temperature probe is connected, it can not be
manually set, and the probe must be closed before setting it
manually.
Convert
temperature
Setting the measured resistance value needs to be converted to the
temperature value. The converted temperature value is 0
℃ ~
+255
℃
, which is related to the accuracy of the converted resistance
value.
Winding
materials
The winding materials of the test products can be selected as copper
and aluminum. The winding materials are related to the conversion
coefficient used in the conversion value of the resistance.
Testing current
Select the test current gear. Choose 1mA, 10mA, 0.1a, 1A, 5A, 10A
and automatic.
2
The secondary operation directory corresponds to the setting
parameters of the primary operation directory, and the parameters
are modified by “←”→” key to move the cursor and “↑↓” key.
Tip: the cursor in the second level of operation directory, you can
press the "Enter" or "Cancel" key to jump the cursor to the first level
of operation directory.
No.
Instruction
3
The range of current selected for testing.
4