7
The voltage V
Current A
A road
00.00
000.0
B road
00.00
000.0
C road
00.00
000.0
When the power indicator of the host machine is on, press the output button on the
voltage regulator. The output button indicates that the red light is on and the voltage
regulator is in the output state. Then adjust the voltage regulator, and there will be
three-phase current output at the large-current terminal of the host. At this point, press
the cut button to stop the current output.
Press the output button on the host, and the output button indicates that the red light
is on. The voltage output at the fixed voltage is 58V. At this point, press the cut button to
stop the voltage output.
VI.Matters needing attention
1. In order to ensure the safety of the test, the device should be reliably grounded before
the test.
2. During the test, it is strictly prohibited to connect and dismantle the test line.
3. The company's supporting input and output lines shall be used.
4. The cross section of the input power cable of the tester shall be greater than 2mm
2
,
large current output connection cable section should be as short as possible output
connection cable length.
5. Prolonged continuous output should be avoided.