21
Choose in sequence
ADJ
,
HC CALC
、
CO CALC
、
CO
2
CALC
、
HC
(
at this time press
▲
and then C
3
H
8
appears
)
and
ESC
via S.Change HC/ C
3
H
8
into C
3
H
8
via S and ▲Change
the concentration of C
3
H
8
, CO and CO
2
according to staItion gas value labeled on bottles
via S and ▲/▼. Inject staition gas for about 5 second through gas inlet; after data shown
in “
Current”
is stable, choose
ADJ
via S and press E to confirm ADJ results. N. Scale
obtained shall be between 0.5 and 1.5.
If N. Scale obtained exceeds the range, “wrong adjust” appears on the right bottom of the
screen. Choose
ESC
via S and return to adjust menu via E.
6.2 O
2
/NO ADJ
Select2”2: O
2
/NO ADJ “in adjust menu via S. Press E and the screen will display
Choose in sequence
O
2
,
NOx NO CALC
and
ESC
via S
6.2.1 O
2
ADJ
Choose
O
2
via S, press E and the screen will show:
HC/CO/CO
2
ADJ
hh : mm
:
ss
Gsuge
Current
New scale
H3C8
1600
0
1.002
CO
3.60
0.00 1.031
CO
2
13.00
0.35 1.001
S :select
,
E :confirm
,∧∨
: modify:
ESC
ADJ
O
2
/NO ADJ
hh
:
mm
:
ss
Gauge
Current
O
2
:
20.90
21.58
NO
:
1030
0
S :select
,
E : modify
,∧∨
:modify
ESC
O
2
NOx