18
(1)
Press
until the display shows “SET COLD JUNC”.
(2)
Enter the desired cold junction temperature by pressing numeric keys and press
, the display shows “COLD JUNC OK”.
(3)
Press any key to exit the “SET COLD JUNC” mode.
VI.Switching Temperature Unit
To switch temperature unit, proceed as follows:
Take switching °C to °F for example:
(1)
Press
until the display shows “SET COLD JUNC”.
(2)
Press
until the display shows “°C TO°F”.
(3)
Using the same way to switch °F to °C.
(4)
Press
.
VII.Calibration
EDEKA TC-200 Multi-function process calibrator can only be calibrated by old hand.
CAUTION: The connections of calibration are the same as connections of using these parameters.
CAUTION: Clear offset of resistance or RTD before using to eliminate offset errors and test leads errors,
according to ‘8.4 Clear of Resistance and RTD’.
7.1 Equipments Required
High-precision multimeter
High- Precision resistance box
Summary of Contents for EDEKA TC-200