7
1:the set value of temperature or time is not
allowed to alter
Parameter table 2
Parameter
prompt
Name
Instruction of the function
(Setting
range)
factory
set value
Lc-
Password key
When Lc=9, enter the next parameters.
0
ndA-
Temp alarm
mode
0: With over-temp alarm only.
1: With over-temp alarm and
under-temp alarm at the same time.
(0
~
1)
0
ndt-
Timer mode
0: No timer function.
1: The timer get to work as soon as the set
temperature value is achieved..
2: The timer start to work as soon as the
instrument get to work.
(0
~
2)
1
Hn-
Timer unit
0: Minute.
1: Hour.
(0
~
1)
0
SPD
Timer
parameter
When measured value of
temperature >SPD+set value of
temperature,timer get to work
(0.1
~
100.0
℃
)
0.5
SPT
Constant
temperature
tip time
In timing mode(set value of temperature is
achieved),the buzzer reminds you when the
tip time you’ve set is achieved.
(0
~
9999S)
0
EST
Tip after timing
When timing program is over,the buzzer
reminds you as soon as the tip time you’ve
set is achieved
(0
~
9999S)
0
EH-
Timer end
0: Continue to maintain the constant
(0
~
1)