Attention! For the sake of safety, when test has

been finished, please remove the meter away

from the lead on test.

3.PEAK Test

PEAK test: the maximum current test. The meter will make automatic comparison on the

changes in measured current, indicating the maximum value and maintain it, when the meter is
removed away from the lead on test, the test result will be kept consistently, suitable for the circuit
test on those LCD data hard to be read directly.

Under the normal testing mode, press PEAK button

PEAK indication light will be bright,

enter

PEAK testing mode. That means the meter will display and automatically keep the maximum value in
the test.

Under any other mode, it must be returned to normal testing mode, and then make PEAK testing

according to the above operation.

Press POWER button to exit from PEAK testing mode, returning to normal testing mode, and it

will display “End” symbol during the exit.

4.Data Maintenance

Under normal testing mode, press HOLD button to keep LCD display, “HOLD” symbol displays.

Press HOLD button again to release data lock, returning to normal testing mode, and “HOLD”
symbol disappears.

5.Data Memory

Under normal testing mode, press HOLD to storage data, at the same time, Meter will

automatically form serial numbers, and memory the current preserved data. During the storage,
“MEM” symbol will show once by flash. This meter can memory 99 groups of data, in case of full
storage, “FULL” symbol will display by flash continuously, and it cannot continue to memory until
cleaning out the previous memory.

6.Data Reference

Under normal testing mode, press PEAK + POWER button to enter data reference mode,

displaying “MR” symbol, meanwhile, it will automatically display data saved in the group 01. Then
press PEAK or POWER button to read the stored data by turning upwards or downwards in cycle. It
will return to the group 1’s data automatically after reaching the last group’s data in storage.

Press HOLD button to exit from data reference mode, returning to normal testing mode. During

the exit, it will display “End” symbol.

Summary of Contents for HZRC 9000

Page 1: ...HZRC9000B High Low Voltage Clamp Meter USER MANUAL Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...CONTENT Warning 1 I Brief Introduction 3 II Electrical Symbols 3 III Series Model 4 IV Technical Specification 4 VI LCD Display 6 VII Operation 8 VIII Replacing Batteries 13 IX Accessories 14...

Page 3: ...strictly forbidden to use this meter to test the wire without any insulate protection has voltage exceed 35KV or convergence generatrix Please pay attention to marked words and symbols on panel or ba...

Page 4: ...eration according to the indication The extremely dangerous symbol the user must make safe operation according to the indication It suggests that this meter shall be made insulation intensity test at...

Page 5: ...reservation data storage wireless transmission and other functions and its special clamp meter can make it easier for clamping or evacuating the tested wire through pressing or pulling back the insula...

Page 6: ...supervision Power Supply DC6V Alkaline Dry Battery 1 5V AAA X 4 Test Method Clamp CT Integral Method Transmission Method B Model Wireless Transmission the maximum distance for direct transmission is...

Page 7: ...lay then press HOLD Key to cancel this function Data Access MR symbol indicates it can read the stored data by turning upwards and downwards Overflow Display Exceed measurement range overflow function...

Page 8: ...Type II V Structure 1 Double input toroid including the boot area 2 HZCR9000 Tester 3 LCD Display 4 PEAK Test Indication 5 PEAK Key 6 POWER Key 7 HOLD Key 8 HZCR9000B Receiver 9 Insulation Rod Connec...

Page 9: ...ol when the storage memory has been filled with 99 groups of data it will display FULL symbol by flash which means it cannot continue to store data MR data access symbol which is displayed during look...

Page 10: ...o damage before usage According to manual instructions to install the battery A Tester Operation 1 Power On Off Press POWER button to power on the instrument will displays the LCD screen entering norm...

Page 11: ...ng personal damage or equipment damage General Test In the process of test LCD will display the tested current value real time and LCD data will vary with the current change When the tester is removed...

Page 12: ...domain perpendicular to the lead when moving away Under HOLD mode press HOLD button to cancel HOLD function and return to normal testing mode Under data reference mode press POWER button for more than...

Page 13: ...intenance Under normal testing mode press HOLD button to keep LCD display HOLD symbol displays Press HOLD button again to release data lock returning to normal testing mode and HOLD symbol disappears...

Page 14: ...ormal testing mode If LCD display is relatively dark after power on it is possible that battery voltage is a bit lower Please replace battery 15 minutes after powering on the meter if it remains idle...

Page 15: ...group 1 s data automatically after reaching the last group s data in storage Press POWER button for more than 3 seconds to exit from data reference mode returning to receiving data mode During the ex...

Page 16: ...ton to check whether the meter can be powered on normally If not please repeat the operation according the step 2 IX Accessories Tester 1PCS Receiver HZRC9000B special 1PCS Insulation Rod 1M Section 5...

Reviews: