3.2 Auto linear increasing mA output
Auto lope current output allow the calibrator to output passive (supply current)
or active (simulate) loop current that is continuously changing. And you can still
leave your hands free to test calibrator responsibility.
Press to enter into “Auto lope current output” model and step to next slope
type.
The calibrator applies or controls a continuously repeated mA signal in one of
below three slope types within the span of 0-20 mA or 4-20 mA:
Slow slope: gentle slope with rate of change 0%-100%-0% within 40s.
Fast slope: abrupt slope with rate of change 0%-100%-0% within 15s.
Stepping slop: 25% stepping slope with 5s of pause per step.
Press any key or switch off the calibrator to exit this function.
Use “Spancheck” function
“Spancheck” function can be used to check the zero point or span of a calibrator
under SOURCE or SIMULATE model.
To select “Spancheck”, please press .