8
2
.4 Properties Description
Property
Text
Description
Name
English Letters or Numbers
Name of the logger
SN
10 Characters
Must be the same as the one in the label of
logger.
Sampling Interval (s)
Number from 1 to 240
Sampling frequency in LCD screen
Logging Interval (s)
Number from 2 to 86400
Logging frequency / S400 Upload frequency
Down Mode
(Cannot be changed)
Start Mode
(Cannot be changed)
Delay Time
(Cannot be changed)
Stop Mode
First In First Out
When recording memory is full, the logger
will delete the earliest data
Stop When Full
When recording memory is full, the logger
will stop record new data
Logs
The count of records in the memory.
Total
The total capacity of the logger’s storage.
Alarm
Turn ON / OFF
the Data Logger's Over Limit Alarm
LCD
Please choose ON
Power
Please choose Normal
High & Low Limit
CH1: Temperature
CH2: Humidity / Temperature2
When the value over the limit, the buzz will
sound / red LED will flash an alarm.
Offset
6
Input positive number to decrease the value.
Input negative number to increase the value.
6
Calibrate the logger: The logger is factory calibrated to an accuracy given in the device specifications. However,
there may be times when you wish to adjust the calibration of your logger. Logpro provides you with the ability to
perform a single point offset calibration. This calibration can be used to increase the accuracy of the logger for a
restricted data range.