GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
The Symmetry IQ 4000 Series is intended for use in the electromagnetic environment specified below. The customer
or the user of the Symmetry IQ 4000 Series should assure that it is used in such an environment.
Immunity Test
IEC60601 test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic discharge
(ESD) IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
supply lines
±1kV for input/output
±2kV for power
supply lines
±1kV for input/output
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC61000-4-11
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
Mains power quality should be
that of a typical commercial
or hospital environment. If
the user of the Symmetry
IQ 4000 Series requires
continued operation during
power mains interruptions,
it is recommended that
the Symmetry IQ 4000
Series be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.