1. Pull straight up on the irrigation bottle to remove the bottle from
the control unit.
2. Fill a spare bottle with at least 175ml with clear distilled water.
3. Place the cap on the bottle, check the irrigating bottle connector
opening and surrounding area of the control unit and clean as
necessary. After cleaning, install the irrigation bottle by presssing
down until it clicks into place. (Fig 16)
CAUTION:
Use only distilled water for cleaning and confirm the
connector hole and surrounding area is clean and free of debris.
4. Turn the irrigation bottle adjustment knob (green) to maximum
water flow. (Fig 17)
5. Select the Irrigation Bottle water selection mode on the display panel.
The Irrigation Bottle lamp will illuminate. (Fig 18)
6. Press and hold the water selection mode key for 2 seconds. The unit
will beep, all the lamps except the irrigation bottle lamp will shut off
and the auto cleaning sequence will begin. The auto cleaning sequence
takes approximately 30 seconds.
7. After the auto cleaning sequence begins, push the water selection
mode again to stop the sequence
CAUTION:
If you press the mode selection key and the power key during the auto cleaning sequence,
the tip will vibrate. Please confirm the power setting does not exceed the maximum recommended
power for the tip.
NOTICE:
Setting the irrigation bottle water adjustment knob at less than maximum during auto cleaning
will result in incomplete cleaning in the 30-second cycle.
8. When the auto-cleaning sequence completes, the unit will revert to the settings prior to cleaning.
9. Remove the irrigation bottle by pulling straight up.
10. After use please thoroughly clean and dry all bottles prior to next use.
manUal cleaning methoD – cleaning oF irrigation tUbe
(Use of Irrigation Bottle)
1. Pull straight up on the irrigation bottle to remove it from the control unit.
2. Open the cap of the irrigation bottle, remove any remaining solution and fill the bottle with distilled water.
3. Close the cap firmly and replace the bottle on the control unit. (Make sure the air hole is clean and free
from debris). Insert the bottle into the irrigation connector opening; push down until it clicks.
4. Operate the unit for approximately 30 seconds with irrigation bottle water flow at maximum setting.
Water line maintenance
Local, regional and federal authorities establish safe potable water quality level guidelines for human use. In
Europe the maximum accepted level of colony forming units of bacteria (cfu’s) for potable water used in
DUWLs is no more than 100 cfu/ml at a temperature of 22°C (20 cfu/ml at a temperature of 36°C). In the
U.S., the Centers for Disease Control established a 500 cfu/ml level, while the American Dental Association
sets a maximum level of 200 cfu/ml. Follow regulatory recommendations in your region regarding flushing
dental unit water lines.
MIN
MAX
W
M
Dry
POWER
17
18