PV-ISOTEST
EN - 35
10.2. GENERAL
CHARACTERISTICS
10.2.1. Reference gudelines
Instrument safety:
IEC/EN61010-1, IEC/EN61010-2-030,
IEC/EN61010-2-033,
IEC/EN61010-2-034
EMC: IEC/EN61326-1
Safety of measuring accessories:
IEC/EN61010-031
General: IEC/EN62446-1
Measurement M
: IEC/EN61557-2
Measurement RPE:
IEC/EN61557-4
Insulation:
double insulation
Pollution level:
2
Measurement category:
CAT III 1500V DC, CAT III 1000V AC,
Max 1500VDC, 1000V AC between inputs
10.2.2. Display and memory
Type of display:
graphic COG 128x128 pxl, with backlighting
Saved data:
max 999 tests
PC interface:
optical/USB
10.2.3. Power supply
Battery type:
6x1.5V alkaline type AA LR06 or
6x1.2V NiMH rechargeable batteries type AA LR06
Low battery indication:
symbol “ ” on the display
Battery duration:
> 500 test (for each function)
Auto power off:
after 5 minutes' idling
10.2.4. Mechanical characteristics
Size (L x W x H)
235 x 165 x 75mm (9 x 6 x 3in)
Weight (batteries included):
1.2kg (2.5lv)
Mechanical protection:
IP40
10.3. ENVIRONMENTAL
CONDITIONS FOR USE
Reference temperature:
23°C ± 5°C (73°F ± 41°F)
Operating temperature:
0°C ÷ 40°C (32°F ÷ 104°F)
Allowable relative humidity:
<80%RH
Storage temperature:
-10°C ÷ 60°C (14°F ÷ 140°F)
Storage humidity:
<80%RH
Max operating altitude:
2000m (6562ft)
This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU
(LVD) and of EMC Directive 2014/30/EU
This instrument satisfies the requirements of European Directive 2011/65/EU (RoHS)
and 2012/19/EU (WEEE).
10.4. ACCESSORIES
See the attached packing list.