background image

  

HTFLEX33D 

 

EN – 5

 

3. INSTALLATION 

  The current probe was designed to allow the operator to connect this measurement 

device around a conductors without disconnecting it as many CTs presently require. 
Even though the current probes output is AC, there are instances where the user will 
want to orient the transducer so that proper polarity will exist at the output terminals. 
This is done by installing the transducers around the conductors with the molded-in 
arrow on the latch (see Fig. 2) pointing in the direction of conventional current flow. 
Conventional current flow is defined as current flowing from the positive to the negative 
potential 

 

 When a three phase plant is under test, the correspondence between volt probe of 

measuring instrument connected and clamp measuring the same phase must be 
respected. Please refer to phase labels (I1, I2, I3) present on flexible clamps and 
interconnection cables 

 

  The current probes must be installed with the interconnection cable on the outside of 

the loop when the latch is engaged. The polarity arrow, the double insulation, and the 
CAUTION symbols will all be on the outside of the loop. It should also be noted that the 
current probes would produce twice the output voltage if you wrap the transducers 
around the conductors twice 

 

  There is minimal shock hazard using a HTFLEX33D current probes. Each transducer 

has been Hi-Pot tested to several thousand volts with no voltage breakdown. This 
particular characteristic allows high-current measurement (with a wide frequency 
bandwidth) of conductors at less than 1000VAC potential to earth. 

 
 

CAUTION 

 

 

 

  Hazardous potentials may exist in the vicinity of the desired current 

measurements. Use locally approved safety procedures when working near 
hazardous potentials 

 

It is recommended not to install the clamp around a live bus that is at a 
hazardous potential

. If installation is not possible when the bus is inactive or 

power is turned off, always use appropriate gloves and/or equipment approved 
for working around hazardous potentials when installing the HTFLEX33D in the 
vicinity of these potentials. 

 

4. MAINTENANCE 

 

CAUTION 

 

 

 

  Remove the detergent with clean water, then wipe dry with a clean cloth. The 

use of solvents as cleaners are not recommended unless thoroughly tested 
and found harmless to all surface and parts. Do not submerge the 
HTFLEX33D transducers or the electronics package into water or other fluids 

  Preventive maintenance primarily consists of cleaning the transducers and 

cable to prevent surface contamination. Use a mild detergent and water to 
clean the transducers and cables. Remove the detergent with clean water, 
then wipe dry with a clean cloth 

 

5. END OF LIFE 

 

CAUTION

: the symbol on the instrument indicates that the appliance and its 

accessories must be collected separately and correctly disposed of 

Summary of Contents for HTFLEX33D

Page 1: ... Copyright HT ITALIA 2013 Release EN 2 00 24 06 2013 ENGLISH User manual ...

Page 2: ...CRIPTION 3 2 1 Introduction 3 2 2 Transducer s functions 4 3 INSTALLATION 5 4 MAINTENANCE 5 5 END OF LIFE 5 6 TECHNICAL SPECIFICATIONS 6 6 1 Reference standards 6 6 2 Technical features 6 6 3 Mechanical features 6 6 4 Environment conditions 6 7 SERVICE 7 7 1 Warranty conditions 7 7 2 Service 7 ...

Page 3: ...e the current flowing is greater than the maximum measurable current Overrange CAUTION Hazardous potentials may exist in the vicinity of the desired current measurements Use locally approved safety procedures when working near hazardous potentials It is recommended not to install the clamp around a live bus that is at a hazardous potential If installation is not possible when the bus is inactive o...

Page 4: ...ring instrument take the clamp off the cable wait 1 minute before connecting the clamp to the measuring instrument then clamp the cable again Do not use the HTFLEX33D if is or seems damaged 2 INSTRUMENT DESCRIPTION 2 1 Introduction HTFLEX33D is an innovative current transducer based on the Rogowsky s principle that combines easy use with measurement accuracy The HTFLEX33D current probe is similar ...

Page 5: ...s from other current measuring methods The measuring transducers are constructed from non ferrous materials minimizing any circuit loading due to magnetic influence Fig 2 Measurement transducer The frequency response of the current probes is rather wide compared to conventional CTs This allows the user to monitor a much wider range of line harmonic components than conventional CTs allow The HTFLEX...

Page 6: ...ing a HTFLEX33D current probes Each transducer has been Hi Pot tested to several thousand volts with no voltage breakdown This particular characteristic allows high current measurement with a wide frequency bandwidth of conductors at less than 1000VAC potential to earth CAUTION Hazardous potentials may exist in the vicinity of the desired current measurements Use locally approved safety procedures...

Page 7: ...Range 3dB 10Hz 20kHz Phase Error 45 65 Hz 1 Position Sensitivity 2 rdg Temperature Coefficient 0 05 rdg per C Working voltage 1000V DC ACrms 6 3 Mechanical features Output connection 2 m cable with 3 way Hypertac connector Length of flexible clamp 610mm 24in Maximum conductor diameter 174mm 7in Weight 215g 8 ounces Material TPE Polypropylene UL94 VO rated Mechanical protection IP65 6 4 Environment...

Page 8: ...be necessary as a consequence of service actions carried out by unauthorized personnel Any modification of the instrument carried out without the authorization of the manufacturer Use not provided for in the instrument s specifications or in the instruction manual The content of this manual cannot be reproduced in any form whatsoever without prior authorization of the manufacturer All our products...

Reviews: