5.8. DC CURRENT MEASUREMENT
CAUTION
The maximum input for DC current is 10A (
10A
input) or 400mA
(
mAµAV
Hz%
input).. Do not attempt to measure higher currents
to avoid electrical shocks or damages to the instrument.
Fig. 9: Using the instrument for DC Current measurement
1. Power off the circuit under test
2. Switch on “
A
”, “
mA
” o “
10A
”
3. Press the
MODE
key until the “DC” symbol is shown at display
4. Insert the test leads into the jacks, the red plug into
10A
or
mAµAV
Hz%
jack and black plug into
COM
jack
5. Connect the red and the black plugs in series with the circuit whose current is to be
measured respecting the polarities (see Fig. 9)
6. Energize the circuit under test. The current value will be displayed.
7. The message "
O.L
" means that the detected current exceeds the limits.
8. The symbol "
-
" on the instrument display indicates that voltage has opposite direction
with regard to the connection.
9. For use of HOLD, RANGE, MAX MIN functions please refer to § 4.2