![background image](http://html1.mh-extra.com/html/ht/f3000u/f3000u_user-manual_2178131038.webp)
NEPTUNE
EN - 36
Reference guidelines
Instrument safety:
IEC/EN61010-1,IEC/EN61010-2-030,
IEC/EN61010-2-033
EMC: IEC/EN
61326-1
M
test:
IEC/EN 61557-2
Lo
test:
IEC/EN 61557-4
Phase sequence:
IEC/EN 61557-7
Insulation:
double insulation
Pollution level:
2
Measurement category:
CAT IV 600V, CAT III 1000V to earth and
between inputs
7.1.1. General
characteristics
Mechanical characteristics
Size (L x W x H):
175 x 85 x 55mm (7 x 3 x 2in)
Weight (batteries included):
420g (15ounces)
Mechanical protection:
IP40
Power supply
Battery type:
4x1.5V batteries type AAA IEC LR03
Low battery indication:
symbol
“”
”
on the display
Battery life:
V, A,
,
approx. 132h (backlight OFF)
V,
A,
,
approx. 68h (backlight ON)
M
(@500V)
approx. 400 tests (backlight
OFF).
Lo
approx. 2000 tests (backlight OFF).
Auto Power Off:
after 15 minutes
’
idling (can be disabled)
Display
Type of display:
4 dgt LCD, max 9999 dots, decimal sign, point
backlight and bargraph, indication of polarity
Sampling frequency:
2 times/s
Conversion: RMS
7.2. ENVIRONMENTAL
CONDITIONS FOR USE
Reference temperature:
23°C
5°C (73°F ± 41°F)
Operating temperature:
5°C ÷ 40°C (41°F ÷ 104°F)
Allowable relative humidity:
<80%RH
Storage temperature:
-20°C ÷ 60°C (-4°F ÷ 140°F)
Storage humidity:
<80%RH
Max operating altitude:
2000m (6562ft)
This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU
(LVD) and of EMC Directive 2014/30/EU.
This instrument satisfies the requirements of European Directive 2011/65/EU (RoHS)
and 2012/19/EU (WEEE).
7.3. ACCESSORIES
See the attached packing list.