EQUITEST
5071
EN - 32
Using modes "P-P", "P-
N", if the instrument
detects an impedance
higher than 199.9
, the
following screen
appears on the display.
LOOP 30.11.12
Symbol ">" indicates that
the value of impedance is
higher than the maximum
measurable value
>199.9
---A
FRQ =50.0HZ
VP-N= 1V VP-PE= 0V
P-N
FUNC Z2
The results can be saved by pressing the SAVE key twice (see § 7.1).
Using modes "P-PE",
"R
A
", if the instrument
detects an impedance
higher than 1999
, the
following screen
appears on the display.
LOOP
30.11.12
Symbol ">" indicates that
the value of impedance is
higher than the maximum
measurable value
>1999
---A
FRQ =50.0HZ
VP-N= 1V VP-PE= 0V
R
A
50V
FUNC
UL
The results can be saved by pressing the SAVE key twice (see § 7.1).