EMC Declaration
24/25
Recommended separation distances between portable and mobile RF
communications equipment and the blood pressure monitor
The
is intended for use in an
electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the
can help
prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters)
and the
as recommended below
according
to the maximum output power of the communications equipment.
“
”
“
”
,
blood pressure monitor
blood pressure monitor
blood pressure monitor
Rated maximum
output power of
transmitter
W
eparation distance according to frequency of transmitter
m
150 kHz to 80 MHZ
80 MHz to 800 MHZ
800 MHz to 2,5 Ghz
0,01
0,1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.