HRM U80K Instruction Manual Download Page 22

Electromagnetic
environment

guidance

Compliance
level

IEC 60601
test level

Immunity test

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 kHz to 80
MHZ

3 V m
80 MHz to 2,5
Ghz

/

3 V

3 V/m

Portable and mobile RF
communications equipment
should be used no closer to any part
of the
including cables

than the

recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.

Recommended separation distance

d=1.2

P

d=1.2

P 80MHz to 800MHz

d=2.3

P 800MHz to 2.5 Ghz

where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
should be less than the compliance
level in each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol:

”,

,

,

.

blood pressure monitor

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the

is used exceeds the applicable RF compliance level above,

the

should be observed to verify normal operation. If abnormal

performance is observed, additional measures may be necessary, such as reorienting or
relocating the

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1
V m

”.

]

/ .

blood pressure monitor

blood pressure monitor

blood pressure monitor

a

b

a

b

EMC Declaration

22/25

Summary of Contents for U80K

Page 1: ...Model U80K Fully Automatic Upper Arm Style Blood Pressure Monitor Instruction manual...

Page 2: ...ors of wrong measurement Fitting the cuff Measuring procedure Discontinuing a measurement Memory recall of measurements Read memory record Memory clear of measurements About blood pressure Exceptional...

Page 3: ...essure and pulse the components are included the body cuff and printed instruction manual Batteries are optional This unit is intended for the adult using Intelligent inflation will reduce the uncomfo...

Page 4: ...dispose this product as unsorted municipal waste Collection of such waste separately for special treatment is necessary Direct current Operating instructions Follow instructions for use Consult accomp...

Page 5: ...is modified appropriate inspection and testing must be conducted to ensure continued safe use of equipment The cuff hose around neck may cause the suffocation The swallowing of samll park like packin...

Page 6: ...ous substances can leak into the groundwater and get into the food chain damaging your health and well being 1 Internally powered equipment 2 Type B applied part 3 Protection against ingress of water...

Page 7: ...cture Only provided cuff can be used can not change to any other branded cuff Product structure 7 25 Body Display Cuff size and connection Year Month Time Date Systolic blood pressure Diastolic blood...

Page 8: ...When power on the low battery symbol will display once the unit start to work and you must replace with new batteries otherwise the unit can t work Please use 4pcs AA identical 1 5V alkaline batteries...

Page 9: ...quirements If in doubt consult your local representative or the technical service department 2 When using AC power to avoid possible damage to the monitor use only the exclusive AC adapter that can be...

Page 10: ...20XX the last digit of the year will increase 1 when press button each time and decrease 1 when press button each time you could choose from 2013 to 2099 Press button SET when you confirm the year th...

Page 11: ...ou confirm the date then it will enter into the time setting mode Continue to above step the screen will display xxMxxD and xx xx and keep flashing on the digits of hour the digit will increase 1 when...

Page 12: ...their arm can increase blood pressure Make sure you are in a comfortable relax position and do not activate any of the muscles in the measurement arm during measurement Use a cushion for support if n...

Page 13: ...bber tube leaves the cuff on the inner side of the arm 4 Tighten the free end of the cuff and close the cuff by affixing the velcro 5 The cuff should be snug on your upper arm so That you can fit 2 fi...

Page 14: ...flash you can hear the beep for every heartbeat once the heartbeat signal is detected 4 When the measurement has been completed you can hear a long beep in the meantime the systolic diastolic and pul...

Page 15: ...ermanently remove all stored memories Press the button SET for 6 times until CL appears when power off press the START STOP button CL will flash for 3 times to clear all the memories After this press...

Page 16: ...re mmHg 110 100 90 85 80 120 130 140 160 180 Systolic bl od pressure mmHg o Optimal blood pressure target value Normal blood pressure Mild hypertension Moderate hypertension Severe hypertension High n...

Page 17: ...Error indicators The following symbol will appear on the display when measuring abnormal Exceptional Situation 17 25 Trouble removal...

Page 18: ...h or detergent mix until it is fully saturated Squeeze any excess moisture from the cloth to avoid any dripping or potential oversaturation of the cuff Wipe all surfaces of the blood pressure monitor...

Page 19: ...Temperature 5 40 Storage environment Temperature 55 Humidity 10 85 avoid crash sun burn or rain during transportation 20 Accuracy Measuring principle Automatic upper arm blood pressure monitor Automa...

Page 20: ...cceptable level The unit is guaranteed to be free of defects in workmanship and materials under normal use for a period of Two Years from the date listed on the purchase record For repair under this w...

Page 21: ...on mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supp...

Page 22: ...marked with the following symbol blood pressure monitor NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propaga...

Page 23: ...CISPR 11 Group 1 The uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment blood pressure m...

Page 24: ...ed maximum output power of transmitter W eparation distance according to frequency of transmitter m 150 kHz to 80 MHZ 80 MHz to 800 MHZ 800 MHz to 2 5 Ghz 0 01 0 1 1 10 100 0 12 0 38 1 2 3 8 12 0 12 0...

Page 25: ...Rev 01 Model U80K Fully Automatic Upper Arm Style Blood Pressure Monitor...

Reviews: