12
HT130003 / HT130042
HT130030 / HT130043
User Manual
CorTemp
®
Data Recorder v4.3
9001 Series
3. Press
F2
to
CLEAR DATA
. The following screen will display:
4. Press
F1
to clear data from recorder. To cancel, press
F2
.
5. Press
ENTER
to return to
EXIT / STBY
mode.
A.14. Adding Sensors
To program the recorder with the HQI Temperature Sensors, do the following:
1. From
the
EXIT / STBY
mode, press
MENU
.
2. Press
READ
until the
ADD SENSOR / SENSOR ACT
screen displays.
3. Press
F1
to
ADD SENSOR
. The following screen will display:
4.
Enter the 2-digit Sensor # (Player #, 01 – 99).
5.
When the Sensor # is entered, the following screen will display:
6. Press
F1
to program the Serial #.
7.
Enter the 6-digit Serial # from the sensor package label.
8. Press
F2
to program the Calibration #.
9.
Enter the 8-digit Calibration # from the sensor package label.
10. When the sensor Serial / Calibration #s have been entered, press
READ
to enter the
Serial / Calibration #s for the next sequential sensor. If no other sensors require
programming, press
ENTER
to return to the
ADD SENSOR / SENSOR ACT
screen.
11. Press
ENTER
to return to
EXIT / STBY
mode.
Sensor Serial / Calibration #s may be programmed into the recorder via the CorTrack
®
II
Software.
NOTE: If any of the calibration numbers have been entered incorrectly, temperature
readings for that sensor will be erroneous. Generally, bad temperature readings will
be so far off as to be immediately obvious. However, there are times when they are
CorTemp
®
CorTemp
®
CorTemp
®
CorTemp
®