Appendix C
HP E1346A Multiplexer Error Messages
Table C-1 lists the error messages associated with the multiplexer modules
programmed by SCPI. See the appropriate mainframe manual for a
complete list of error messages.
Table C-1. 48-Channel Multiplexer Error Messages
No.
Title
Potential Cause(s)
-211
Trigger ignored
Trigger received when scan not enabled. Trigger received after scan complete.
Trigger too fast.
-213
Init Ignored
Attempting to execute an INIT command when a scan is already in progress.
-224
Illegal parameter value
Attempting to execute a command with a parameter not applicable to the command.
1500
External trigger source
already allocated
Assigning an external trigger source to a switchbox when the trigger source has
already been assigned to another switchbox.
2000
Invalid card number
Addressing a module (card) in a switchbox that is not part of the switchbox.
2001
Invalid channel number
Attempting to address a channel of a module in a switchbox that is not supported
by the module (e.g., channel 99 of a multiplexer module).
2006
Command not
supported
on this card
Sending a command to a module (card) in a switchbox that is unsupported by the
module.
2008
Scan list not initialized
Executing a scan without the INIT command.
2009
Too many channels in
channel list
Attempting to address more channels than available in the switchbox.
2012
Invalid Channel Range
Invalid channel(s) specified in SCAN <
channel_list>command. Attempting to begin
scanning when no valid channel list is defined.
2600
Function not supported
on this card
Sending a command to a module (card) in a switchbox that is not supported by the
module or switchbox.
2601
Channel list required
Sending a command requiring a channel list without the channel list.
Appendix C
HP E1346A Multiplexer Error Messages 71