5-44
Hardkey and Softkey Reference
FM
FM Stop Rate
Use this softkey to change the ending internal modulation frequency for
swept-sine frequency modulation. To change the value, press the softkey once.
The current value for the FM stop rate is displayed in the active entry area. To
enter a new value, rotate the front panel knob until the desired value is displayed,
use the up and down arrow keys, or enter the value using the numeric keypad and
press the
Hz
or
kHz
terminator softkey. The range of values allowed is 0.1 Hz to
50 kHz. The minimum increment allowed is 0.1 Hz.
This softkey is coupled to the
FM Tone 2 Rate
softkey in the FM Dual-Sine Menu.
Any value set for
FM Stop Rate
is reflected in the
FM Tone 2 Rate
softkey.
Conversely, if you change the value for the
FM Tone 2 Rate
softkey, the value for
FM Stop Rate
is changed to match.
Softkey Location: Press
FM/
Φ
M
,
More (1 of 2)
,
FM Waveform
,
Swept-Sine
,
FM Stop Rate
Status after Normal Preset: 400.0 Hz
SCPI Commands:
[:SOURce]:FM[1]|2:INTernal[1]:FREQuency:ALTernate <val><unit>
[:SOURce]:FM[1]|2:INTernal[1]:FREQuency:ALTernate?
FM Sweep Time
Press this softkey to set the sweep time for swept-sine frequency modulation. The
signal generator will vary the FM rate from the specified start rate to the stop rate
in the time set with this softkey. Enter the desired sweep time using the up and
down arrow keys, the front panel knob, or enter a value using the numeric keypad
and press the
sec
or
msec
terminator softkey. The range of values allowed is 1 ms
through 65.535 s. The minimum increment allowed is 1 ms.
Softkey Location: Press
FM/
Φ
M
,
More (1 of 2)
,
FM Waveform
,
Swept-Sine
,
FM Sweep Time
Status after Normal Preset: 100.0 msec
SCPI Commands:
[:SOURce]:FM[1]|2:INTernal[1]:SWEep:TIME <val><unit>
[:SOURce]:FM[1]|2:INTernal[1]:SWEep:TIME?
Summary of Contents for ESG Series
Page 3: ...ii ...
Page 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Page 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Page 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Page 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Page 111: ...5 2 Hardkey and Softkey Reference ...
Page 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Page 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Page 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Page 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Page 161: ...5 52 Hardkey and Softkey Reference FM ...
Page 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Page 171: ...5 62 Hardkey and Softkey Reference Help ...
Page 173: ...5 64 Hardkey and Softkey Reference Hold ...
Page 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Page 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Page 197: ...5 88 Hardkey and Softkey Reference Local ...
Page 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Page 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Page 201: ...5 92 Hardkey and Softkey Reference Mode ...
Page 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Page 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Page 235: ...5 126 Hardkey and Softkey Reference Preset ...
Page 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Page 245: ...5 136 Hardkey and Softkey Reference Recall ...
Page 247: ...5 138 Hardkey and Softkey Reference Return ...
Page 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Page 255: ...5 146 Hardkey and Softkey Reference Save ...
Page 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Page 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Page 312: ...7 1 7 Operation This chapter is reserved for future use ...
Page 313: ...7 2 Operation ...
Page 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...
Page 341: ...9 18 Specifications HP ESG Series Signal Generators Specifications ...