33
8)
If the connection between the PMIS program and the AT200
Interface board is lost the error when starting the PMIS
program will appear: ‘
Can not find the AT200 camera
controller’
. The first step is then to power down the
computer, wait for about 30 seconds and then start the
computer again. The PMIS program is then likely to find the
AT200 Interface board.
If the problem still exists the reset button on the AT200
camera controller might have to be pushed when the starting
the computer. This reset button is located on the top of the
interface board, (the computer must be opened to be able to
press the reset button), see AT200 CCD Camera System,
Hardware reference manual and the appropriate World Wide Web
pages from Photometrics and GKR Computer Consulting.
9)
If the HP SureStore tape bacup drive is not responding and
gives the following error message:
1.
Shutdown and restart computer
2.
Run ScanDisk and Defrag
3.
Close other applications
4.
Try a different tape
5.
Uninstall Colorado Backup, and reinstall
then the most likely reason is that there is a bad electrical
connection to the backup drive. Open the computer, deconnect
and reconnect the power connector to the tapedrive, and do the
same with the SCSI connector (both ends).
Please take any precaution to prevent Electro Static Discharge,
ESD. The air on Svalbard is very dry, so the risk of ESD when
working inside the computer is considerable. THIS CAN DAMAGE
THE WHOLE COMPUTER!
(Espen 99/01/15)