Chapter 2
2-15
Quick Start: Learning How to Make Measurements
Learning to Make Reflection Measurements
Step 2. Make a measurement calibration
Follow these instructions to perform an S
11
1-port error correction:
1. Select a calibration kit that is appropriate to your device under test. Press
. Choose the calibration kit that is appropriate to your
test device by pressing the appropriate softkey. For example, if your test device uses
type-N 50
Ω
connectors, press
. If your test device uses 7-mm connectors, press
, and so on.
2. Press
twice,
.
3. Follow the prompts shown on the analyzer display to connect and measure an open,
short, and load on PORT 1.
Any choice of male/female in the calibration process should always be made for the sex
that represents the test port. For example, if the test port had a male, type-N connector,
you would connect the female, type-N calibration device. But when you follow the
prompts on the analyzer to measure a short calibration standard, you would select
, or the sex that represents the test port.
CAUTION
To ensure an accurate error correction, you must connect the calibration
standards to the adapters or cables that you will include in the actual device
measurement.
NOTE
If a mistake is made, standards can be measured more than once before
pressing
. Only the last measurement data is used.
Figure 2-9
Connections for an S
11
1-Port Error-Correction
4. Press
after measuring the three standards.
5. Press
.
6. Press
to complete the process.
Cal
CAL KIT SELECT CAL KIT
N 50
Ω
7mm
RETURN
CALIBRATE MENU S11 1-PORT
SHORT (M)
DONE 1-PORT CAL
DONE 1-PORT CAL
Save/Recall
SAVE STATE