Chapter 2
2-7
Quick Start: Learning How to Make Measurements
Learning to Make Transmission Measurements
Step 2. Perform a measurement calibration
1. Disconnect your test device from the analyzer.
2. Connect a “thru” between the measurement cables, as shown in
Figure 2-3
. Include all
the adapters that you will use in your device measurement.
If noise reduction techniques are needed for the measurement, the instrument's
settings (reduced IF BW, and /or averaging) should be selected prior to any
error-correction.
Figure 2-3 Connections for a “Thru” Calibration Standard
3. Press the following keys to make a transmission response calibration:
4. To save the error-correction (measurement calibration), press:
5. Next, choose from the following options:
• Choose
if you want to save the calibration results and
instrument state to the analyzer's memory.
• Choose
if you want to save the calibration results and
instrument state to the disk that is in the analyzer's internal disk drive.
• Choose
if you want to save the calibration results and
instrument state to the disk that is in an (optional) external disk drive that is
configured to the analyzer.
6. Press
to save the error-correction (measurement calibration).
NOTE
Example procedures for all types of error-correction (measurement
calibrations) are located in the “Calibrating For Increased Measurement
Accuracy” chapter in the User's Guide. For information on the analyzer
operation during error-correction (measurement calibration), refer to the
“Operating Concepts” chapter in the User's Guide.
Cal
CALIBRATE MENU RESPONSE THRU
Save/Recall
SELECT DISK
INTERNAL MEMORY
INTERNAL DISK
EXTERNAL DISK
RETURN SAVE STATE