9
System performance summaries
HP 8753ET (50-ohm systems)
Type-N test ports
Configuration
Network analyzer
HP 8753ET Standard,
Option 006 or Option 004
Calibration kit
HP 85032B
Test port cable
HP part number 8120-4781
System dynamic range
300 kHz to 16 MHz
100 dB
16 MHz to 3 GHz
110 dB
3 GHz to 6 GHz
105 dB
Measurement port characteristics
Frequency range
Uncorrected
300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHz
Directivity
30 dB
24 dB
19 dB
Source match
4
25 dB
20 dB
14 dB
Load match
24 dB
19 dB
16 dB
Reflection tracking
±1.0 dB
±1.0 dB
±2.0 dB
Transmission tracking
±1.5 dB
±1.5 dB
±2.5 dB
Crosstalk
100 dB
100 dB
90 dB
Frequency range (continued)
Corrected
300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.0 GHz
Reflection measurements
1
Directivity
50 dB
47 dB
40 dB
Source match
42 dB
36 dB
31 dB
Load match
One-port cal
24 dB
19 dB
16 dB
Enhanced reflection cal
24 dB
19 dB
16 dB
Reflection tracking
±(0.009 dB
±(0.019 dB
±(0.07 dB
+0.01 dB/˚C)
+0.02 dB/˚C)
+0.03 dB/˚C)
Transmission measurements
2
Source match
Enhanced-response cal
42 dB
36 dB
31 dB
Response-only cal
3
25 dB
20 dB
14 dB
Transmission tracking
Enhanced-response cal
±(0.006 dB
±(0.018 dB
±(0.054 dB
+0.01 dB/˚C)
+0.02 dB/˚C)
+0.03 dB/˚C)
Response-only cal
±(0.033 dB
±(0.1 dB
±(0.27 dB
+0.01 dB/˚C)
+0.02 dB/˚C)
+0.03 dB/˚C)
1. One-port or enhanced-response calibration.
2. Enhanced-response or response-only calibration.
3. 2 dB less with Option 004.
4. Option 004 may degrade uncorrected source match as much as 2 dB.
Transmission uncertainty:
(enhanced-response calibration)
Transmission uncertainty:
(response calibration)
Reflection uncertainty
(one-port calibration)