6
System performance summaries
HP 8753ES (50-ohm systems)
3.5-mm test ports
Configuration
Network analyzer
HP 8753ES
Standard and Option 006
Calibration kit
HP 85033D
Test-port cables
HP 11857D
System dynamic range
30 kHz to 50 kHz
70 dB
1
50 kHz to 300 kHz
90 dB
1
300 kHz to 16 MHz
100 dB
16 MHz to 3 GHz
110 dB
Option 014
108 dB
3 GHz to 6 GHz
105 dB
Option 014
103 dB
Reflection uncertainty
Magnitude
Transmission uncertainty
Magnitude
Measurement port characteristics
Frequency range
Corrected
30 kHz-300 kHz
2
300 kHz-1.3 GHz 1.3 GHz-3 GHz
3 GHz-6 GHz
Directivity
49 dB
46 dB
44 dB
38 dB
Source match
49 dB
44 dB
41 dB
37 dB
Load match
49 dB
46 dB
44 dB
38 dB
Reflection
±(0.010 dB ±(0.005 dB
±(0.007 dB
±(0.009 dB
tracking
+0.02/˚C) +0.01/˚C)
+0.02/˚C)
+0.03/˚C)
Transmission
±(0.016 dB ±(0.014 dB
±(0.022 dB
±(0.048 dB
tracking
+0.02/˚C) +0.01/˚C)
+0.02/˚C)
+0.03/˚C)
Phase
Phase
1. Typical below 300 kHz.
2. Typical performance.