26
Verification kits
Measuring known devices other than the standards
used in calibration is an easy way to verify the proper
operation of an HP 8753ES measurement system. HP
offers verification kits which include devices, with
data, for verifying the error-corrected measurements
of an HP 8753ES and 50-ohm test sets.
HP 85029B 7-mm verification kit
Contains a set of precision 7-mm devices, with data
traceable to NIST* for verifying the calibrated perfor-
mance of an HP 8753E measurement system.
The HP 85031B 7-mm calibration kit is required for
complete verification.
Test-port return cables
Hewlett-Packard offers high-quality RF cables to con-
nect the HP 8753ET/ES and test sets to devices under
test. These cables offer excellent RF shielding for
high- dynamic-range measurements.
HP 11851B 50-ohm type-N RF cable kit
Recommended for use with the HP 11850C/D three-
way power splitters. Kit includes three phase-matched
610-mm (24-in.) cables and one 860-mm (34-in.)
cable.
Return loss
> 24 dB to 3 GHz
Phase tracking
±4˚ at 1.3 GHz
HP 8120-5639 50-ohm type-N RF cable
A 50-ohm type-N test port cable (both connectors are
type-N male).
Return loss
> 30dB to 6 GHz
HP 11857B 75-ohm type-N test port
return cables
A pair of 610-mm (24-in.) test port return cables for
use with the HP 8753ES or HP 85046B 75-ohm S-para-
meter test set.
Return loss
> 24 dB to 2 GHz
Phase tracking
±2˚ at 1.3 GHz
HP 11857D 7-mm test-port return cables
A pair of 610-mm (24-in.) test port return cables for
use with the HP 8753ES with the HP 85046A and
85047A S-parameter test sets. These cables can be
used with connector types other than 7-mm with the
appropriate precision adapters.
Return loss
> 24 dB to 3 GHz
> 20 dB to 6 GHz
Phase tracking
±2˚ at 1.3 GHz
HP 11857F type-F test-port return cables
Type-F test port cable set (75-ohm). Includes one
cable with type-N(male) to type-F(male) connectors
and one cable with type-N(male) to type-F(female)
connectors.
HP 11850C/D three-way power splitters
HP 11850C
HP 11850D
Impedance
50 ohms
75 ohms
Frequency range
DC to 3 GHz
DC to 2 GHz
Tracking
±25 dB, ±3˚
±2 dB, ±2.5˚
Equivalent
source match
30 dB at 1.3 GHz 30 dB at 1.3 GHz
(ratio or leveling) 20 dB at 3 GHz
20 dB at 2 GHz
Nominal
9.5 dB + 1 dB/GHz 7.8 dB
insertion loss
Input port match
DC to 1.3 GHz
20 dB
20 dB
1.3 GHz to F
max
10 dB
10 dB
Maximum
operating level
+20 dB
+20 dB
Damage level
+30 dB
+30 dB
RF connectors
RF input 50-ohm type-N (f) 50-ohm type-N(f)
All others 50-ohm type-N (f) 75-ohm type-N(f)
Includes
Three HP 11852B 50-to 75-ohm
minimum-loss pads
Recommended accessories
HP 11851B RF cable kit
HP 11667A 50-ohm power splitter
Frequency range
DC to 18 GHz
Typical insertion loss
6 dB
Equivalent source match
DC to 4 GHz
26 dB
4 GHz to 8 GHz
21 dB
8 GHz to 18 GHz
17 dB
Tracking
(between output arms)
DC to 4 GHz
±15 dB
4 GHz to 8 GHz
±2 dB
8 GHz to 18 GHz
±25 dB
Maximum operating level
±27 dBm
Connectors
50-ohm type-N (f)
Option 001
RF input
type-N (m)
Outputs
type-N (f)
Option 002
RF input
type-N (f) on RF
Outputs
precision 7-mm
Dimensions
46 mm H x 52 mmW x
19 mm D
(1.8 x 2.0 x 0.7 in)
Recommended accessories
HP 11851B RF cable kit
HP 11852B 50-ohm to 75-ohm minimum loss pad
Frequency range
DC to 3.0 GHz
Nominal insertion loss
5.7 dB
Return loss
300 kHz to 2 GHz
32 dB
2 GHz to 3 GHz
27 dB
Maximum input power
250 mW (+24 dBm)