1-19
General Information
Care of Fiber-Optic Connectors
If the performance, after the initial cleaning, seems poor try cleaning the con-
nector again. Often a second cleaning will restore proper performance. The
second cleaning should be more arduous with a scrubbing action.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good indica-
tion that you have good connections. Since return loss and insertion loss are
key factors in determining optical performance they can be used to determine
connector degradation A smooth, polished fiber end should produce a good
return loss measurement. The quality of the polish establishes the difference
between the “PC” (physical contact) and the “Super PC” connectors. Most
connectors today are physical contact which make glass-to-glass connections,
therefore it is critical that the area around the glass core be clean and free of
scratches. Although the major area of a connector, excluding the glass, may
show scratches and wear, of the glass has maintained its polished smoothness,
the connector can still provide a good low level return loss connection.
If you test your cables and accessories for insertion loss and return loss upon
receipt, and retain the measured data for comparison, in the future you will be
able to tell if any degradation has occurred. Typical values are less than 0.5 dB
of loss, and sometimes as little as 0.1 dB of loss with high performance con-
nectors. Return loss is a measure of reflection: the less reflections the better.
The larger the return loss, the smaller the reflection. The most physically con-
tacting connectors have return losses better than 50 dB, although 30 to 40 dB
is more common.
Summary of Contents for 86060C Series
Page 1: ...User s Guide HP 86060C Series Lightwave Switches ...
Page 18: ...1 12 General Information Specifications and Regulatory Information ...
Page 46: ...3 6 Using the Switch Changing Switch Position Changing Switch Position ...
Page 93: ...5 Spare Channel Replacement Procedure 5 4 Electrostatic Discharge Information 5 7 Servicing ...
Page 104: ...Index 4 Index ...