Assembly Replacement Procedures
Test Port Connector
Replacement
2.4 mm Test Port Connectors
1. Incline the rear of the test set approximately four inches and support it
with a stable, solid object (such as a thick book).
2. Using a 9/16-inch open-end wrench, unscrew the test port connector nut.
3. Pull out the center conductor assembly and replace it with a new
assembly. Do not touch either end of the new assembly. Part numbers for
the test port connector components are located on page 7-16.
4. Replace the test port connector nut and torque it to 562 N-cm (50 in-lb).
5. Measure the pin depth of each port using a 2.4 mm female pin-depth
gage. Refer to the connector care information in the calibration kit
manual for more information about using the gage.
NOTE
If the center pin protrudes, or if the depth is less than 0.0001 inch, remove
the connector assembly and increase or decrease shims to adjust the pin
depth to between 0.0001 and 0.0007 inch. When measuring the pin depth be
sure to include the uncertainty calculation of your gage.
6. If you added or removed shims, regage the test port connector assembly.
Figure 6-3
Diagram of 2.4 mm Test Port Connector
Summary of Contents for 8517B
Page 10: ...x HP 8517B S Parameter Test Set Manual ...
Page 16: ...Contents 6 HP 8517B S Parameter Test Set Manual ...
Page 30: ...1 14 HP 8517B S Parameter Test Set Manual Getting Started Operating and Safety Precautions ...
Page 38: ...2 8 HP 8517B S Parameter Test Set Manual Installation Packaging the Test Set ...
Page 57: ...HP 8517B S Parameter Test Set Manual 1 ...
Page 60: ...HP 8517B S Parameter Test Set Manual 4 4 Specifications ...