4-3
HP 8517B S-Parameter Test Set Manual
Specifications
Rear Panel RF Connectors
Connector type
precision 2.4 mm female
Recommended connection torque
90 N-cm (8 in-lb), maximum for precision-2.4 mm hardware
Damage input level
15 dBm
1
Nominal connector-nut size
8 mm
User 1 Power Levels for Reference Channel Phase Lock
Minimum power level
–45 dBm
Maximum power level
–10 dBm to –20 dBm, depending on frequency
HP 8517B Power Requirements and Physical Characteristics
Operating temperature range
0
°
C to 55
°
C
Input Power (AC)
110, 120, 220, or 240 Vac, ±10%
Line frequency from 47 to 66 Hz
Dimensions
460 mm (High)
×
133 mm (Wide)
×
609 mm (Deep)
(18.1 inches H
×
5.25 inches W
×
24 inches D)
Weight
15 kg (35 lb), net