HP 8517B S-Parameter Test Set Manual
3-13
Making Operational Checks
To check the operation of multiple test-set configurations, do the following:
1. Connect a device with a known response to test set number 1. Press the
following HP 8510C front-panel keys:
INSTRUMENT STATE
[LOCAL]
AUXILIARY MENUS
[SYSTEM]
{HP-IB ADDRESSES}
{TEST SET}
2. When the message
TEST SET HP-IB ADDRESS
appears, enter
[20]
, the
address of test set number 1, then press
[x1]
. The measurement trace from
test-set number 1 should display.
3. Store the trace in memory by pressing the following keys:
MENU
[DISPLAY]
{DATA AND MEMORIES}
{DATA --> MEMORY n}
DISPLAY
: {DATA and MEMORY]
4. Next, enter the HP-IB address of test set number 2, followed by
[x1]
, then
re-enter the address of test set number 1 followed by
[x1]
.
5. Repeat the above procedure for each of the other test sets in the system.
Observe any difference in the response between the stored trace and the
results trace by switching back and forth between the test sets.
Any differences in data that you suspect are due to the IF switch (Option
001) or to RF switching, must be checked out. Refer to Chapter 5,
“Troubleshooting the Test Set” in this manual.
Summary of Contents for 8517B
Page 10: ...x HP 8517B S Parameter Test Set Manual ...
Page 16: ...Contents 6 HP 8517B S Parameter Test Set Manual ...
Page 30: ...1 14 HP 8517B S Parameter Test Set Manual Getting Started Operating and Safety Precautions ...
Page 38: ...2 8 HP 8517B S Parameter Test Set Manual Installation Packaging the Test Set ...
Page 57: ...HP 8517B S Parameter Test Set Manual 1 ...
Page 60: ...HP 8517B S Parameter Test Set Manual 4 4 Specifications ...