Error Messages
If an error condition occurs in the sweeper, it will always be reported to both
the front panel and HP-IB error queues. These two queues are viewed and
managed separately. The
key is used to view the contents of the front
panel error queue. The HP-IB query
is used to view the
contents of the
error queue.
If there are any error messages in the front panel error queue, the front panel
MSG annunciator will be lit. Pressing the
key repeatedly until the MSG
annunciator turns off will empty the front panel error queue. The
key
has no affect on the HP-IB error queue. Emptying the
error queue
has no affect on the front panel queue, therefore, it will not affect the MSG
annunciator.
There are some special error types that are called permanent errors.
Permanent errors remain in the error queues until the error condition is
cleared. Pressing the
key will empty the front panel error queue, but
the permanent errors will be re-reported if the error conditions still exist.
In the HP-IB error queue, the permanent errors are re-reported after the
message,
error” is read using the
query or after the
command is executed.
16-2
Summary of Contents for 83751A
Page 1: ...User s Guide HP 83751A B and HP 83752AlB Synthesized Sweepers ...
Page 22: ...Contents 17 3 Typical Phase Noise 10 GHz Carrier 17 12 17 4 Dimensions 17 17 Contents 8 ...
Page 24: ...1 Performing the Operator s Check ...
Page 35: ...I 2 Externally Leveling the Sweeper ...
Page 46: ...3 Generating a Stepped Sweep ...
Page 48: ...4 Generating a Millimeter Signal ...
Page 54: ...5 Creating User Flatness Arrays I ...
Page 59: ......
Page 63: ...I I 6 Operating a Master Slave Setup ...
Page 70: ...I 7 Front Rear Panel ...
Page 100: ...I 8 Instrument State Keys ...
Page 112: ...I 9 Marker Keys ...
Page 121: ...10 Modulation Keys ...
Page 128: ...11 Frequency Keys ...
Page 141: ...12 Sweep Keys ...
Page 151: ...13 Power Keys ...
Page 161: ...14 Entry Keys ...
Page 162: ...Entry Keys pg48ab Figure 14 1 The Entry Group 14 2 ...
Page 171: ...I 15 Special Functions ...
Page 172: ......
Page 199: ...16 Error Messages ...
Page 235: ...I 17 Specifications and Options I ...
Page 244: ...Specifications and Options RF output Source Match 1 7 1 SWR internaLLy Leveled 1 7 1 0 ...
Page 245: ......
Page 252: ...Specifications and Options General Weight Net 16 kg 35 lb Shipping 23 kg 49 lb 17 18 ...
Page 257: ...18 Safety and Regulatory Information ...
Page 266: ...Index ...