Distribution
Amplier
T
est
5.
Set
the
synthesized-source
frequency
to
300
MHz
and
adjust
its
amplitude
for
a
04.0
dBm
reading
on
the
power
meter
.
Save
these
synthesized-source
settings
in
Register
5
by
pressing
4
STORE
5
4
5
5
on
the
synthesized-source
front
panel.
6.
Set
the
synthesized-source
frequency
to
300
MHz
and
adjust
its
amplitude
for
a
+4.0
dBm
reading
on
the
power
meter
.
Save
these
synthesized-source
settings
in
Register
6
by
pressing
4
STORE
5
4
6
5
on
the
synthesized-source
front
panel.
7.
Connect
the
synthesized
source
to
A
IN
on
the
device
under
test
(DUT).
8.
One
by
one
,
connect
the
power
sensor
to
the
DUT's
A
1
OUT
,
A
2
OUT
,
and
A
3
OUT.
Each
output
should
have
a
power
level
of
0
dBm
61.0
dB
(01.0
dBm
to
+1.0
dBm).
9.
Press
4
RECALL
5
4
5
5
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
given
in
step
5.
Then
repeat
step
8.
10.
Press
4
RECALL
5
4
4
5
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
given
in
step
4.
11.
One
by
one
,
connect
the
power
sensor
to
the
DUT's
A
1
OUT
,
A
2
OUT
,
and
A
3
OUT
.
Each
output
should
have
a
power
level
of
0
dBm
62.0
dB
(02.0
dBm
to
+2.0
dBm).
12.
Press
4
RECALL
5
4
3
5
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
given
in
step
3.
Repeat
step
11.
13.
Press
4
RECALL
5
4
2
5
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
given
in
step
2.
Repeat
step
11.
14.
Press
4
RECALL
5
4
1
5
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
given
in
step
1.
Repeat
step
11.
15.
Press
4
RECALL
5
4
6
5
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
given
in
step
6.
16.
Repeat
steps
7
through
14
for
the
DUT's
distribution
amplier
\B".
4-14