![HP 5004A Operating And Service Manual Download Page 28](http://html.mh-extra.com/html/hp/5004a/5004a_operating-and-service-manual_161471028.webp)
4-8. NEGATIVE PULSE PERFORMANCE 'TEST
4-9. With test equipment connected as in Figure 4-3, proceed as follows:
a.
Set pulse generator to output a negative-going pulse.
b.
Adjust pulse generator to give waveform at probe tip as shown in Figure 4-3, with a
repetition rate of one-pulse-per-second. Probe indicator should flash off approximately
once per second.
c.
Disconnect test equipment.
61
04A
POWER
SUPPLY
\
PULSE
\
I
GENERATOR
I
LEADS SHORT
Figure
4-3.
Negative Pulse Performance Test Setup
4-10.
DATA PROBE SETUP 'TIME PERFORMANCE TEST
4-11. Connect the equipment as shown in Figure 4-4. Equipment front panel settings:
80138 Front Panel Settings:
Pulse period
=
200 ns (5 MHz) in 20 n position
Pulse width
=
square wave
Amplitude
=
5V.
NOTE
Adjust the
pulse width to obtain approximately the same
pulse period of
throughout the frequency range.
80078
Front Panel Settings:
External Input
-
Ext. Trigger
Pulse delay
-
5.0 ns position
Pulse width
-
5.0 ns position
Slope Polarity
+
Transition time
-
2.0 ns Leading edge: Fully CCW. Trailing edge: Fully CCW
-
NORM
Amplitude
=
+5V
Output Pulse Polarity
+
Summary of Contents for 5004A
Page 1: ......
Page 9: ......
Page 18: ......
Page 20: ......
Page 22: ......
Page 24: ......
Page 25: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 33: ......
Page 34: ......
Page 44: ......
Page 55: ......
Page 57: ......
Page 58: ......
Page 61: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......