TB 9-6625-086-35
8
(6) Position the controls of the power meter calibrator as follows:
(a) RANGE (MW) switch to .01.
(b) FUNCTION switch to CAL.
NOTE
After each change in position of TI RANGE switch, turn the
FUNCTION switch of power meter calibrator to the CURRENT
OFF position and adjust the TI zero controls for a zero
indication.
(7) Turn the TI and power meter calibrator RANGE switches to the settings listed
in table 4. If the indications are not within limits specified, perform
b
below.
Table 4. Range and Trucking Performance Check (with Power Meter Calibrator)
Test Instrument
Power Meter
Calibrator
Test Instrument Indications (mw)
(Readings for Model 431B
are in parenthesis)
Adjustments for Model
RANGE switch
RANGE (MW)
Min
Max
200-Ohm
100-Ohm
settings
switch settings
Mount
Mount
.01
.01
0.0098
(0.0097)
0.0102
(0.0103)
R14
R1
.03
.03
0.0295
(0.0291)
0.0305
(0.0309)
R13
R2
.1
.1
0.099
(0.097)
0.101
(0.103)
R12
R3
.3
.3
0.297
(0.291)
0.303
(0.309)
R11
R4
1
1
0.990
(0.970)
1.010
(1.030)
R10
R5
3
3
2.970
(2.910)
3.030
(3.090)
R9
R6
10
10
9.900
(9.700)
10.100
(10.300)
R8
R7
10
1
8
7.9
(7.7)
8.1
(8.3)
10
6
5.9
(5.7)
6.1
(6.3)
10
4
3.9
(3.7)
4.1
(4.3)
10
2
1.9
(1.7)
2.1
(2.3)